6-18
6 SPECIAL MODE (Menu Mode)
6.4 Setting Mode 1
In this mode, the following settings are possible.
1. Setting minimum reading
2. Auto power off
3. Vertical and horizontal angle tilt correction (Tilt ON/OFF)
4. Systematic error of instrument correction
5. Setting for RS-232C communication
6. Selecting communication port (only for
Bluetooth
function built-in model)
7. Confirming the
Bluetooth
device address (only for
Bluetooth
function built-in
model)
8. Setting humidity input
9. Setting NP-TRK mode
10. Setting EDM eco. mode
11. Setting volume
•
This setting is memorized after power off.
6.4.1 Setting Minimum Reading
Select minimum display unit for angle measurement, coarse distance measurement mode.
To select minimum display for fine measurement mode, see Chapter 17 “SELECTING MODE” .
[Example] Minimum angle: 5”, Coarse: 1 mm
Models
Angle Unit
Coarse mode
Distance unit
Degree
GON
MIL
GM-52/55
1" / 5" 0.0002 gon / 0.0010 gon 0.005 mil / 0.02 mil
Prism mode, Non-Prism mode
10 mm (0.02 ft)/1 mm (0.005 ft)
Operating procedure
Operation
Display
1
After pressing the
{MENU}
key, press the
{F4}
(P
↓
) key twice to get the menu on page 3.
{MENU}
{F4}
{F4}
2
Press the
{F2}
key.
{F2}
3
Press the
{F1}
key.
{F1}
4
Press the
{F1}
key.
{F1}
5
Press the
{F2}
(5") key and press the
{F4}
(ENTER) key.
{F2}
{F4}
MENU 3/4
F1:ILLUMINATION
F2:PARAMETERS 1
F3:PARAMETERS 2 P
↓
PARAMETERS 1 1/4
F1:MINIMUM READING
F2:AUTO POWER OFF
F3:TILT P
↓
MINIMUM READING
F1:ANGLE
F2:COARSE
MINIMUM ANGLE
[F1: 1" ]
F2: 5"
ENTER
MINIMUM READING
F1:ANGLE
F2:COARSE