2-3-3 Trouble shooting (Main body)
No.
Trouble
Cause
Remedy
Tool
Reference
page
1
The slit does not
close.
Dust is stuck to the
slit.
Wipe off dust with
ether many times.
•
Eyepiece with
scale
P.22, 23
The slit is wedge-
shaped.
Adjust the slit blade.
•
Eyepiece with
scale
P.22, 23
2
There is a differ-
ence between the
aperture of
φ
0.2 and
the slit center.
The accuracy is not
good in the incorpo-
ration of the slit unit
and illumination
unit.
Loosen the slit box
tightening screws (4
pcs.) to position
them accurately.
•
Standard scale
•
Eyepiece with
scale
P.16
3
Difference between
the slit center and
microscope center.
The reflection mir-
ror in the illumina-
tion unit is not
positioned correctly.
Position the reflec-
tion mirror in the
illumination unit
accurately.
•
Standard scale
•
Eyepiece with
scale
P.17, 18
R-SLD7-0-0308-11