60
BEFORE REQUESTING SERVICE
"Fail to execute chart software. (Error Code
602)”
This message is displayed when the MC-4S starts and
the necessary files cannot be accessed. Try the follow-
ing items.
Restart the CV system.
Reinstall the software.
If the instrument does not reset, ask the TOPCON office
to repair it.
"Could NOT find the selected chart monitor.
Please confirm the chart monitor setting and
the connection."
or
"Fail to execute chart software. (Error Code
603)"
It is probable that the monitor cable is not connected or
is connected improperly. Connect the cable and restart
the CV system.
It is probable that the MC-4S is connected through a
VGA cable. Use the DVI cable to connect with the MC-
4S. After connecting, restart the CV system.
It is probable that the connected monitor is not the MC-
4S. Make sure that the MC-4S is connected as the
monitor. Make sure that "MC-4S" is set in "Chart moni-
tor selection". After connecting, restart the CV system.
"Fail to execute chart software. (Error Code
604)"
It is probable that any other chart except the MC-4S is
set in "Chart monitor selection" of "Chart settings".
Set "MC-4S" in "Chart monitor selection" of "Chart set-
tings".
"Fail to execute chart software. (Error Code
605)"
This message is displayed in the following case: When
the CV system starts the MC-4S system, it specifies the
chart type. The MC-4S cannot recognize this setting.
Restart the CV system.
If the instrument does not reset, select a chart type
again or reinstall the software.
"Fail to execute chart software. (Error Code
606)"
This message is displayed when the setting file of the
MC-4S cannot be accessed.
Restart the CV system.
If the instrument does not reset, ask the TOPCON office
to repair it.
"Fail to open chart software configuration
screen because the other task is in progress.
Please try again."
This message is displayed when the command to open
the MC-4S setting screen cannot be executed because
the MC-4S is executing another task.
Access the setting screen again.
"Fail to change chart type because the other
task is in progress. Please try again."
This message is displayed when the command to
change the type cannot be executed because the MC-
4S is executing another task.
Change the type again.
"Fail to change chart language setting
because the other task is in progress."
This message is displayed when the command to
change the language cannot be executed because the
MC-4S is executing another task.
Change the language again.
"Fail to open chart software configuration
screen. (Error Code 300)"
Restart the CV system.
If the instrument does not reset, ask the TOPCON office
to repair it.
"Fail to open chart software configuration
screen. (Error Code 301)"
This message is displayed when the setting file of the
MC-4S cannot be accessed.
Restart the CV system.
If the instrument does not reset, ask the TOPCON office
to repair it.
Error message
Remedial measure