ALADDIN
- User manual
Rev. 19 EN of 10/07/2019
137
For each services the needed parameter are:
Remote Application Entity (AE) title
Remote IP address
Remote connection port
The “Local Application Entity title” is the identifier name through which the device presents itself to the
servers.
The “N-EVENT Report node port” is the port at which the device is able to receive N-EVENT REPORTS for
storage commitments (default is 115).
The connectivity to the defined server can be tested using the “C-ECHO” function activated by the relative
“Connection Test” button. The result of connection test is shown by the green or red icon.
In order to configure properly the full DICOM workflow it could be necessary to perform some operations or
configurations on the server’s side. In order to do this contact the System Administrator.
The DICOM module of ALADDIN HW3.0 is describe in detail in its DICOM Conformance Statement. Visit
http://www.topconmedical.com/conformance.cfm