BD-4601 (BPM Arm 1500)
5.3
Blood Pressure Standard
This device is equipped with a WHO Blood Pressure Indicator which classifies your blood pressure
measurements into six stages (Optimal to Severe hypertension) based on WHO classification on blood
pressure levels as shown in below chart.
After each measurement is completed, the LCD display will show your position automatically on the six
segments of the bar indicator which corresponds to WHO Blood Pressure Indicator
2
.
5.4
Irregular Heartbeat Detector (IHB)
The symbol
will appear on the display, indicating a certain heartbeat irregularity was detected
during measurement. The heartbeat rhythm that is more than or less than 25% from the average rhythm
is usually defined as an irregular heartbeat rhythm. Talking, moving, shaking or an irregular pulse
during the measurement can result in the appearance of this symbol.
Usually this is not a cause for concern, however if the symbol appears often, we recommend you seek
medical advice.And please note that the device does not replace a cardiac examination, but serves to
detect pulse irregularities at an early stage.
Stages of blood pressure levels
Indicator
Systolic
in mmHg
Diastolic
in mmHg
Our recommendation
Grade 3- Severe hypertension
Red
180
110
Take medical treatment at once
Grade 2 - Moderate hypertension Orange
160 ~ 179 100 ~109 Seek medical advice
Grade 1 - Mild hypertension
Yellow
140 ~ 159
90 ~ 99
Check it regularly by doctor
High - normal
Green 3
130 ~ 139
85 ~ 89
Check it regularly by doctor
Normal
Green 2
120 ~ 129
80 ~ 84
Check it yourself
Optimal
Green 1
< 120
< 80
Check it yourself
•
No Irregular Heartbeat Detector measurement will be stored in memory.
•
The pulse display is not suitable for checking the frequency of heart
pacemarkes. If a certain pulse irregularity is detected during measurement
often, we recommend you seek medical advice.
•
However, as a safeguard, we recommend that if you have arrhythmias such
as atrial or ventricular premature beats and atrial fibrillation or any other
special conditions you should check with your physician before using your
device.
•
The IHB function is not designed for use by people with arrhythmias nor for
diagnosing or treating an arrhythmic problem. In order to filter the unstable
status of user and avoid affecting the detection of heart rate from any
movement, shaking or talking in the beginning of measurement, the method
of averaging heart beat intervals for subject device is calculated with the
three proper heart beat pulses detected in the beginning of measurement
and that is different from a strict mathematical averaging of all recorded
intervals.
•
At least 3 beats with at least 25% difference from the average heart beat
interval will generate the IHB icon on the screen.