6.5. Replacing Peristaltic Pump Tubes
WARNING
After rinsing with deionized water, reagent tubes may still
contain traces of chemicals
Wear appropriate protective equipment (gloves, glasses,
protective clothing).
1
Flush the reagent tubes with deionized water as described in
Stop of Operation for Maintenance, p. 32
2
Turn the power switch to the OFF position.
3
Loosen the three screws [A] on the front of the pump and remove
the front cover [B].
4
Remove the tube from the rotor [C] and disconnect it from both
fittings [D].