Cleaning the Control Unit
To clean your RecoveryWrap System, please follow these simple
steps:
1. Remove electrical cord, fold down handle, and remove
top assembly.
fresh, room temperature water.
3. Replace top assembly, electrical cord, and raise handle to lock.
4. Turn system ON for 10 minutes with setting on maximum COLD.
5. Repeat these actions periodically.
RecoveryWrap
Size (approximately)
8”diameter x 11.8”H (20.3cm x 30.0cm)
Weight (dry)
5 lbs. (2.27 kg)
Fluid Tether
8 . (2.4m)
Type
Manual user control
Range
40°F to 70°F (4.4°C to 21.1°C)
Operating Period
2-6 hours typical con nuous use
Reservoir Capacity
1.1 gallons (4.1 liters)
Reservoir Fluid
Ice water
Flow Rate (through Vital Wrap)
5.5 gph (21 lph)
Maximum Pressure
18 psi
Voltage
1 V C
Max Power
18W
Leakage Current
Under 300µA
Classi cation
Class II Equipment
Type of Equipment
Type BF
Atmospheric Pressure
525 to 795 mmHg
Humidity
30% to 70% rela ve non-condensing
Temperature
10°C to 40°C
Atmospheric Pressure
179 to 760 mmHg
Humidity
10% to 95%
Temperature
0°C and +50°C
EN60601-1-2(2007)
Regulatory
Operating Environment
Transport & Storage Environment
Regulatory Approvals
Physical
Control System
Thermal System
Circulating System
Electrical System
20 A , 60HZ
Leaks
NOTE: Some condensation(wetness) on the lines and pads is
unavoidable, especially in warmer climates.
1. If during the use of a pad you experience a leak (other than
condensation) disconnect the pad coupling from the
hose by depressing the blue tabs. Reconnect the pad to the hose
2. If the coupling continues to leak or the leak is demonstrated
in the pad itself, STOP using the unit. Contact customer service.
Cleaning the RecoveryWrap Pad:
The RecoveryWrap pad can be washed in cold water using
gently cycle in a conventional laundry machine.
*Top Shelf Orthopedics recommends a maximum of three
machine washes per Recovery wrap.
The Caution or Warning symbol precedes an operational
step that could damage the instrument if the user does not take
certain precautions. Cautions or Warnings are located in the
main text, are preceded by a Caution or Warning statement and
are accompanied by this symbol in the left margin.
Caution: Risk of electric shock.
“OFF” (only for a part of EQUIPMENT)
“ON” (only for a part of EQUIPMENT)
Safety Symbols And Warnings
DANGER –To reduce the risk of electric shock:
1. Always unplug this appliance from the electrical outlet
immediately after using and before cleaning.
2. DO NOT use while bathing or in a shower.
3. DO NOT place or store appliance where it can fall or
be pulled into a tub or sink. DO NOT place in or drop
into water or other liquid.
4. DO NOT reach for a products that has fallen into water.
Unplug immediately.
Periodically check skin where pad is applied for skin
discoloration, numbness, blisters, etc., which can indicate a poten-
tial burn. If detected, discontinue use and consult a physician.
An appliance should never be left unattended when
plugged in. Unplug from outlet when not in use, and before
Close supervision is necessary when this appliance is used
by, on, or near children, invalids, or disabled persons. This appliance
should not be used by or on children without parental supervision.
Use this appliance only for its intended use as described
in this manual. DO NOT use attachments not recommended by
the manufacturer.
Never operate this appliance if it has a damaged cord
or plug, if it is not working properly, if it has been dropped or
damaged, or dropped into water.
Warnings
1. Ensure the top of the appliance is properly installed and the
handle is fully engaged prior to carrying the appliance.
2. DO NOT carry this appliance by supply cord or use cord
as a handle.
3. Keep the cord away from heated surfaces.
4. Never operate the appliance without the Fluid Canister, the Top
Assembly, the Tubing Set, and the pad completely connected.
5. Never operate the appliance with any foreign objects
in the Fluid Canister.
6. Never drop or insert any object into any opening.
remove plug from outlet.
8. Use cooling surfaces carefully. Do not use over insensitive skin
areas or in the presence of poor circulation on an incapacitated
person may be dangerous. Consult your physician.