Top shelf Orthopedics RecoveryWrap System Скачать руководство пользователя страница 2

 

 

 

  

 

1851 East Paradise Road Suite A

Tracy, CA 95304

Phone: 1.866.592.0488

www.topshelforthopedics.com

02419 Rev A

Warranty

Top Shelf Orthopedics guarantees the RecoveryWrap System 

free from defects in material and workmanship for a period of 

6 months from the date of purchase except as noted below. 

This warranty extends only to the original retail purchaser.

This Top Shelf Orthopedics product warranty does not cover 

damage caused by misuse or abuse; accident; the attachment 

of any unauthorized accessory; alteration to the product; or any 

other conditions whatsoever that are beyond the control of

Top Shelf Orthopedics.

Note: For single patient use only.

Cleaning the Control Unit

To clean your RecoveryWrap System, please follow these simple 

steps:

1. Remove electrical cord, fold down handle, and remove 

     top assembly. 

     fresh, room temperature water.

3. Replace top assembly, electrical cord, and raise handle to lock.

4. Turn system ON for 10 minutes with setting on maximum COLD.
5. Repeat these actions periodically.

RecoveryWrap 

         

Size (approximately)

8”diameter x 11.8”H (20.3cm x 30.0cm)

Weight (dry)

5 lbs. (2.27 kg)

Fluid Tether

8  . (2.4m)

Type

Manual user control

Range

40°F to 70°F (4.4°C to 21.1°C)

Operating Period

2-6 hours typical con nuous use

Reservoir Capacity

1.1 gallons (4.1 liters)

Reservoir Fluid

Ice water

Flow Rate (through Vital Wrap)

5.5 gph (21 lph)

Maximum Pressure

18 psi

Voltage

1 V C

Max Power

18W

Leakage Current

Under 300µA

Classi cation

Class II Equipment

Type of Equipment

Type BF

Atmospheric Pressure

525 to 795 mmHg

Humidity

30% to 70% rela ve non-condensing

Temperature

10°C to 40°C

Atmospheric Pressure

179 to 760 mmHg

Humidity

10% to 95%

Temperature

0°C and +50°C

EN60601-1-2(2007)

Regulatory

Operating Environment

Transport & Storage Environment

Regulatory Approvals

Physical

Control System

Thermal System

Circulating System

Electrical System

20 A ,  60HZ

Leaks

NOTE:  Some  condensation(wetness)  on  the  lines  and  pads  is 

unavoidable, especially in warmer climates.

1. If during the use of a pad you experience a leak (other than

 condensation) disconnect the pad coupling from the

    hose by depressing the blue tabs. Reconnect the pad to the hose

2. If the coupling continues to leak or the leak is demonstrated  

     in the pad itself, STOP using the unit. Contact customer service.

 

 Cleaning the RecoveryWrap Pad: 

The RecoveryWrap pad can be washed in cold water using 

gently cycle in a conventional laundry machine.

*Top Shelf Orthopedics recommends a maximum of three 

machine washes per Recovery wrap.

               The Caution or Warning symbol precedes an operational

step that could damage the instrument if the user does not take 

certain  precautions.  Cautions  or Warnings  are  located  in  the 

main text, are preceded by a Caution or Warning statement and 

are accompanied by this symbol in the left margin.

                  Caution: Risk of electric shock.

                 “OFF” (only for a part of EQUIPMENT)

                 “ON” (only for a part of  EQUIPMENT)

Safety Symbols And Warnings

               DANGER –To reduce the risk of electric shock:

1.  Always unplug this appliance from the electrical outlet

     immediately after using and before cleaning. 

2. DO NOT use while bathing or in a shower.

3. DO NOT place or store appliance where it can fall or 

    be pulled into a tub or sink. DO NOT place in or drop 

    into water or other liquid.

4. DO NOT reach for a products that has fallen into water.

    Unplug immediately.

           Periodically check skin where pad is applied  for skin 

discoloration, numbness, blisters, etc., which can indicate a poten-

tial burn. If detected, discontinue use and consult a physician.  

 

                An  appliance  should  never  be  left  unattended  when 

plugged  in.  Unplug  from  outlet  when  not  in  use,  and  before 

                     Close supervision is necessary when this appliance is used 

by, on, or near children, invalids, or disabled persons. This appliance 

should not be used by or on children without parental supervision.

                   Use this appliance only for its intended use as described 

in this manual. DO NOT use attachments not recommended by 

the manufacturer.

                 Never operate this appliance if it has a damaged cord 

or plug, if it is not working properly, if it has been dropped or 

damaged, or dropped into water.

Warnings

1. Ensure the top of the appliance is properly installed and the 

     handle is fully engaged prior to carrying the appliance.

2. DO NOT carry this appliance by supply cord or use cord  

    as a handle.

3. Keep the cord away from heated surfaces.

4. Never operate the appliance without the Fluid Canister,  the Top

    Assembly, the Tubing Set, and the pad completely connected. 

 

5. Never operate the appliance with any foreign objects 

     in the Fluid Canister.

6. Never drop or insert any object into any opening.

     remove plug from outlet.

8. Use cooling surfaces carefully. Do not use over insensitive skin  

      areas or in the presence of poor circulation on an incapacitated

     person  may be dangerous. Consult your physician.

         

NO part of the pad should directly touch the skin.