Humidity:
Normal Working: <90%RH
Reference Working: <80%RH
Transport Environment: ≤93RH
1.4.5 Precautions for Usage
Please do not use the tester in dusty, vibrative, direct sunlight and corrosive gases and other adverse
environments.
When the instrument is not used for a long time, please put it in the original box or similar box and stored in
a ventilated room with temperature of 5
℃
~40
℃
and relative humidity less than 85% RH. Do not store the
tester in a corrosive atmosphere containing harmful impurities and should avoid direct sunlight.
The instrument has been carefully designed to reduce clutter due to AC power input. However, it should still
be used under low noise conditions. If that is inevitable, please install the power filter.
1.4.6 Dimension
W*H*D: 430×185×473 (mm
3
)
1.4.7 Weight
18kg
1.4.8 Safety
This tester is a Class I safe equipment.
1.4.8.1 Insulation resistance
In the reference working conditions, the insulation resistance between the power terminals and the
shell is not less than 50M
.
In hot and humid transport conditions, the insulation resistance between the power terminals and the
shell is not less than 2M
.
1.4.8.2
Dielectric strength
In the reference working conditions, the power terminal and the shell can withstand for one minute
with no breakdown and flashover phenomenon of 1.5kV rated voltage and 50Hz frequency of AC
voltage.
1.4.8.3
Leakage current
The leakage current is not greater than 3.5mA.
1.4.9 Electromagnetic Compatibility
Power Transient Sensitivity, refer to requirements of GB6833.4.
Conducted Susceptibility, refer to requirements of GB6833.6.
Radiated interference, refer to requirements of GB6833.10.