Toho Electronics TRM-10C Скачать руководство пользователя страница 94

93

 

HXPRM10mnCT002E 

Example) 

When calibrating the resistance temperature detector at Channel 2 

 

 

Enter the engineering mode (See the key operation 

in page 78).  Use the 

 key to indicate "

 

", and press the 

ENT

 key. 

Use  the 

  key  to  select  "

",  and  press  the 

ENT

 key. 

Select  the  channel  where  an  instrument  for 

calibration such as a dial resistor is connected, and 

press the 

ENT

 key. 

Enter 100

Ω

.  After 10 seconds, make sure that the 

ALM lamp unilluminated and press the 

ENT

 key. 

Enter 150

Ω

.  After 10 seconds, make sure that the 

ALM lamp unilluminated and press the 

ENT

 key. 

Enter 300

Ω

.  After 10 seconds, make sure that the 

ALM lamp unilluminated and press the 

ENT

 key. 

If the calibration is correct, select "

" with 

the 

  key,  and  if  incorrect,  select  "

", 

respectively.  Then, press the 

ENT

 key. 

 

 

 

 

 

 

 

 

 

 

 

 

 

Display 

Operation Keys 

Description 

 

ENT 

 

 

 

ENT 

 

PRINT 

 

 

 

 

 

PRINT 

 

ENT 

 

 

 

 

ENT 

 

PRINT 

 

ENT 

 

 

 

 

 

ENT 

 

 

 

 

ENT 

 

 

 

ENT 

 

PRINT 

 

 

 

 

Note

 

During calibration, ALM lamp indicates following conditions below. 

Unilluminating the ALM lamp 

 

 

Blinking the ALM lamp 

   

The calibration is within the range. 

   

 

 

 

Judges the input value. 

 

 

 

 

Illuminating the ALM lamp 

 

   

Make sure that 

   

the connection is  

   

properly or input is correctly. 

 

 

 

 

If the calibration value outside the input range was applied , 

(voltage)  or 

(RTD) is displayed when you return to the measurement mode .  

In that case, to calibration again or, please return to the state of the factory default calibration value 

by performing the " Initializing the Setup Data and Calibration Data " described in Section 7.2.8. 

 

Indicates unilluminating the lamp. 

 

Indicates illuminating the lamp. 

RUN   

 

  ALM

 

 

 

CH.No

 

 

RUN     

 

ALM

 

 

 

CH.No

 

 

7

 

RUN   

 

  ALM

 

 

 

CH.No

 

 

6

 

Содержание TRM-10C

Страница 1: ...TRM 10C HYBRID RECORDER PEN TYPE RECORDER INSTRUCTION MANUAL TOHO ELECTRONICS INC HXPRM10mnCT002E SEP 2016 6th Edition Copyright 2000 2016 TOHO ELECTRONICS INC All Rights Reserved...

Страница 2: ...e system is expected by the error occurs due to external factors or failure the instruments take precautions to ensure the overall safety of your system 3 Take precautions to so that an alien substanc...

Страница 3: ...carefully by the instrument 5 When the manual is lost or stained contact our sales representative 6 It is prohibited to copy or reproduce this manual without our permission 1 When installing this inst...

Страница 4: ...autioning information in this manual Symbols used on this manual WARNING Failure to observe this information could result in death or injury Be absolutely certain to read this CAUTION Failure to obser...

Страница 5: ...2 6 4 Digital Print 43 6 4 1 Manual Print 43 6 4 2 List Print 44 6 4 3 Engineering List Print 46 6 4 4 Record Start End Print 48 6 4 5 DI Manual Print option 48 6 4 6 DI Date Time Print option 48 6 4...

Страница 6: ...2 When the Recorder Does not Work atAll 112 10 1 3 When There is a Big Error 113 10 1 4 When the Pen Recording Deflected 113 10 1 5 When Prints Nothing 114 10 1 6 When Printing is Faint 114 10 1 7 Whe...

Страница 7: ...esentative Following accessories should be attached Chart paper Ribbon cassette Cartridge Pen Mounting bracket Instruction manual Packing Fig 1 1 Accessories Table 1 1 List of Accessories Part Name Ty...

Страница 8: ...Open the chart cover to check it This instrument is shipped with the ribbon cassette back of the chart holder Chart Guide Chart Cover Note This instrument is equipped with the ribbon cassette back of...

Страница 9: ...e nameplate is affixed to the middle far side of the instrument Make sure that this instrument meets your requested specification seeing the following tables Table 1 2 Type 01 1 pen type Model 02 2 pe...

Страница 10: ...original nameplate is attached it may damage the door or mounting part CAUTION Storage in a poor environment may damage the appearance functions and service life of the instrument Storage Environment...

Страница 11: ...10 HXPRM10mnCT002E 2 CONSTRUCTION 2 1 Appearance Fig 2 1 Appearance Cartridge pen Door Chart Holder Display keyboard Terminal block Case Mounting bracket Indication card...

Страница 12: ...U V W X Y Z RUN CH No DATA Alarm Type Display Indicates an alarm type in red H is illuminated when the alarm is High And L is illuminated when the alarm is Low Neither of them is illuminated when the...

Страница 13: ...t and change to the setting mode ESC key ESC function Exits that menu halfway selecting a function PRINT key PRINT function Used to perform manual print or list print key function Used to select a set...

Страница 14: ...d safety of the instrument it is recommended to secure the spacing larger than the parenthesized dimensions per unit Unit mm Front Rear Side Panel Cutting 56 144 144 3 5 2 5 100 RUN ALM RUN MENU PRINT...

Страница 15: ...m 2 3 WARNING Do not install the instrument in a place exposed to a combustible explosive or corrosive gas SO2 H2S etc CAUTION Install the instrument in the following places Aplace free from where hu...

Страница 16: ...e IP65 Prior to mounting the instrument to the panel attach packing to the position shown in the figure The rest of the procedure is the same Fig 3 4 Mounting to the Panel in compliance with the IP65...

Страница 17: ...rms to the voltage of the supply source Attach a transparent protective cover prior to turning on the POWER of the instrument Any interruption of the protective conductor inside or outside the apparat...

Страница 18: ...ng conductor there may be an effect caused by noise coming from the grounding conductor It is recommended not to share it with other instrument In order to comply with the requirements of safety stand...

Страница 19: ...ecommended to use a shielded input line For Pt 100 JPt 100 50m max When it is likely to be affected by induction noise particularly when wiring near the high frequency power source it is recommended t...

Страница 20: ...s refer to Fig 4 3 Page 19 Fig 4 4 and Fig 4 5 Page 20 3 Put back the transparent protective cover Fig 4 4 Input Wiring mV V TC CAUTION The transparent protective cover should surely remove the left a...

Страница 21: ...istor to the input terminal block of the instrument Input accuracy is effected with the shunt resistor Use the following recommended resistor Resistance 250 Rated power 1 4W Tolerance 0 1 max Temperat...

Страница 22: ...ive load 0 1Aat maximum L R 7ms at maximum Inductive load Attach an anti surge protective circuit surge absorbers etc to an output terminal as required Attach a round press fitting terminal with insul...

Страница 23: ...iring Example CAUTION The DI Option consists of a combination of 3 Digital inputs The alarm output consists of 3 Relay output Normally open Relay No 1 is a common output of paper empty sensor Option a...

Страница 24: ...4 8 Fig 4 8 Alarm Output Wiring 4 3 3 DI Wiring Procedure Wire the DI refer to Fig 4 9 Fig 4 9 DI Wiring L N 1A 1C 2A 2C 3A 3C 3 Relay output 3C 3A Relay No 3 Normally open L N 1C 1A 2A 3C 3A 2C 1 B 1...

Страница 25: ...igh frequency power source it is recommended to use a shielded twisted wire Attach a round press fitting terminal with insulated sleeve for M3 5 to the end of the electric wire Ground the shield of a...

Страница 26: ...cords the right side approx 85 respectively Fig 5 2 Automatically Moving the 2 Pens and Printer CAUTION It is recommended to use our original chart paper to ensure proper recording If the chart holder...

Страница 27: ...3 Moving up the Pen Push up the pen up lever to move a pen tip away from the chart paper Move up the pen Fig 5 4 Moving up the Pen CAUTION Maximum angle of the door in opened state is 135 degree Do no...

Страница 28: ...to this side Fig 5 5 Taking out the Chart Holder 5 Opening the Chart Cover and Chart Guide Open the chart guide and the chart cover outwardly Fig 5 6 Opening the Chart Cover and Chart Guide Pull out t...

Страница 29: ...sure to loosen the paper Fig 5 7 Loosening the Chart Paper 7 Setting the Chart Paper into the Storage Chamber Unfold the chart paper by two plies Holding the printing surface upward set it in the stor...

Страница 30: ...gning the Chart Paper with Sprocket Drum Chart Holder Sectional View 9 Closing the Chart Cover and the Chart Guide Close the chart cover and the chart guide in the arrow directions Fig 5 10 Closing th...

Страница 31: ...ded to feed the chart paper by 4 plies Fig 5 11 Checking Paper Feed 11 Putting Back the Chart Holder Push in the chart holder horizontally into the case until it is locked Fig 5 12 Putting Back the Ch...

Страница 32: ...ressing the FEED key to Check Paper Feed Press the FEED key on the display keyboard to feed the chart paper Fig 5 13 Pressing the FEED key 14 Pressing the RUN key to Restart Operation UN CH No DATA Th...

Страница 33: ...ock the door Pull the door to this side to open it See Fig 5 3 on Page 26 3 Detaching the Cartridge Pen Pull out the cartridge pen to this side Fig 5 14 Detaching the Cartridge Pen Note When pulling o...

Страница 34: ...aching the Cartridge Pen Reattach the cartridge pen to the pen holder Push it in firmly until it clicks Fig 5 16 Reattaching the Cartridge Pen 6 Pressing the RUN Key to Restart Operation Pen Cap Pen H...

Страница 35: ...up the pen see Fig 5 4 on Page 26 4 Taking out the Chart Holder Put your fingers onto the levers at both sides of the chart holder and pull it out to this side See Fig 5 5 on Page 27 The guide shaft p...

Страница 36: ...9 Detaching from the Main Shaft Viewing from the right side section the hook of the ribbon cassette has been engaged with the main shaft Fig 5 20 Ribbon Cassette Sectional View Ribbon Guide Shaft Lowe...

Страница 37: ...te Pull out the ribbon cassette horizontally If the gear of the ribbon cassette touches the printer move the printer further to the right 9 Unslacking the Ink Ribbon Set the new ink ribbon Turn the ge...

Страница 38: ...n shaft in such a manner that the hook of the ribbon cassette will be engaged with the main shaft Fig 5 24 Reattaching to the Main Shaft Ribbon Cassette Sectional View Main Shaft Guide Shaft Ribbon Ca...

Страница 39: ...ing to the Guide Shaft 13 Putting back the Chart Holder Put back the chart holder horizontally into the case until it is locked See Fig 5 12 on Page 30 14 Moving down the Pen to Close the Door Use the...

Страница 40: ...ay Screen at Power on Note When electric power failure occurs the initialization is performed after power recovery And it becomes the above initial state The printout data are eliminated in initializa...

Страница 41: ...g Colors Channel No Color Red Green 6 2 2 Printing Color Various digital prints are recorded in purple 6 2 3 Pen Gap Fig 6 2 shows the gap between No 1 and No 2 pens Fig 6 2 Gap between Pens 6 2 4 Gap...

Страница 42: ...Recording Operation To start stop recording press the RUN key To stop hold down the RUN key for 3 seconds or more RUN lamp is unilluminated when stop recording 6 3 2 Feed the Chart Paper The chart pa...

Страница 43: ...ge Priority in Reference Note Logging print will not be the wait printing state At the start time of Logging print if the prior Logging print is printing the next one cannot print Prolong the printing...

Страница 44: ...t Note Analog recording is continued while manual print is running Chart paper is fed forcibly If an alarm is activated while manual print is running alarm print will be performed when manual print en...

Страница 45: ...ling value engineering unit Setting alarm type Comment Others 1 List print operating procedure Press the PRINT key Use the PRINT key to display and press the ENT key Use the PRINT key to select Pressi...

Страница 46: ...AG01 NORMAL 0 0 50 0 CH2 mA SCALE 4 00 20 00mA 0 00 100 00 CH LV ALM RLY 1 1 H 28 0 1 1 2 L 18 0 2 4 H 30 00 3 CH DIGITAL ZONE mm 1 ON 0 50 2 OFF 50 100 CH PARTIAL 2 50 40 00 COMMENT 2 TRM PEN 3 12345...

Страница 47: ...he PRINT key to select Pressing the ENT key executes engineering list print Once engineering list print starts the display is automatically returned to the Data Display screen Upon termination of engi...

Страница 48: ...00 2 EXT 805 ON 0 0 0 8000 COM_ADR 01 SPEED 9600BPS DATA 8bit PARITY NONE STOP 1bit DI 1 COMMENT1 2 MAN_PRINT 3 TIMEPRINT Alarm Print Set Value ON 1 occurrence recovery Logging Print Set value ON OFF...

Страница 49: ...l Print option DI manual print prints measurements and time on paper by turning on DI The content of the print is the same as 6 4 1 Manual Print 1 DI function of the engineering mode is set Please ref...

Страница 50: ...to the setup mode 1 DI function of the engineering mode is set Please refer to 7 2 9 DI function for details Date Time print is synchronous print Date Time print is asynchronous print 2 DI is operate...

Страница 51: ...rvals of 2 4 seconds Data renewal period 120ms 6 5 2 Manual Display Indicates the Process variable of specific channel it updates every measurement cycle 120ms The data is the same as Auto display Pre...

Страница 52: ...turn from the setup mode to the recording mode user mode press the MENU key for 3 seconds or more again The following describes how to set the following items The following shows a display map of the...

Страница 53: ...rding Partial compression expansion Digital print Tag Comment words Setting the password Zone Recording Partial Compression Expansion Digital Print Tag Comment Words Setting Mode Setting Copy from Cha...

Страница 54: ...rm OFF relay OFF Engineering unit all channels BF 43 00 Chart speed 20mm h 20mm h Time 2000 01 01 00 00 Sets up the present time Japan standard time GMT 09 00 Zone setting all channels 0 100 Partial c...

Страница 55: ...difference average mode to Unnecessary channels can be skipped mode Setting Channel Mode Key Current Voltage Input Key Thermocouple Input Resistance temperature detector Input Scaling Square Root Deca...

Страница 56: ...n and numeral Use the key to shift a digit Press the ENT key Use the key to select a sign and numeral Use the key to shift a digit Press the ENT key Press the ENT key Setting is completed To return to...

Страница 57: ...56 HXPRM10mnCT002E Note In the or mode press the ENT key to select the range Use the key you can select the type out of the range ENT Key mV mV mV mV Range Mode...

Страница 58: ...the key to select a range 50 mV Press the ENT key Use the key to select a numeral Use the key to shift a digit Press the ENT key Use the key to select a numeral Use the key to shift a digit Press the...

Страница 59: ...T key Use the key to select a numeral Use the key to shift a digit Press the ENT key Press the ENT key Setting is completed To return to the user mode hold down the MENU key for 3 seconds or more FEED...

Страница 60: ...a numeral Use the key to shift a digit Press the ENT key Use the key to select a numeral Use the key to shift a digit Press the key again in the lowest digit right end to blink a decimal point Use the...

Страница 61: ...le L SCALR Scaling upper limit value Scale R IN Input voltage OUT Output Scaling value When an input value is 1 or more 1 to 100 OUT SCALR SCALL SCALL When an input value is less than 1 OUT IN SPAN L...

Страница 62: ...e the key to shift a digit Press the ENT key Use the key to select a numeral Use the key to shift a digit Press the ENT key Use the key to select a numeral Use the key to shift a digit Press the ENT k...

Страница 63: ...ides by the following relational expression OUT1 IN OUT 10 OUT1 LGSCALL Log10 SCALL LGSCALR Log10 SCALR Example When makes setting in previous page the display is as follow table Input voltage V Displ...

Страница 64: ...s the ENT key Setting is completed To return to the user mode hold down the MENU key for 3 seconds or more About Difference Sum and Average The channel to be set must be larger than the one on which c...

Страница 65: ...leted To return to the user mode hold down the MENU key for 3 seconds or more Note Setting all the channels to Skip indicates an error At least one channel should be set to measurement The skip channe...

Страница 66: ...arm set value of 2 000 and alarm output relay No 1 Hold down the MENU key for 3 seconds or more to enter the setup mode Use the key to indicate Press the ENT key Use the key to select the channel you...

Страница 67: ...ress the ENT key If the alarm output is set to ON set the relay number Use the key to select the relay number out of 1 to 3 and press the ENT key Press the ENT key Setting is completed When you want t...

Страница 68: ...character in the same manner When the number of unit characters is less than 6 enter a code to end Press the ENT key Press the ENT key Setting is completed To return to the user mode hold down the ME...

Страница 69: ...68 HXPRM10mnCT002E 1 Character Code Table Example The code number for C is 43...

Страница 70: ...more to enter the setup mode Use the key to indicate and press the ENT key Use the key to select the 1st chart speed and press the ENT key Use the key to set the 1st chart speed and press the ENT key...

Страница 71: ...chart speed Restriction of printing by chart speed When the chart speed is 0 h it s printed with forcible chart feed When the chart speed is more than 120 h not printing as shown the Table above List...

Страница 72: ...nter the setup mode Use the key to indicate and press the ENT key Use the key to select the copy from channel Use the key to select the copy to channel and press the ENT key Press the ENT key To retur...

Страница 73: ...ther the measured data for each channel is to be printed or not Tag Set the tag which is to be printed instead of a channel number for logging print or manual print Setting up to 5 characters See Page...

Страница 74: ...nting position in percentage Use the key to select a numeral and use the key to shift a digit Press the ENT key Set a span printing position in percentage Operate in the same manner as setting the zer...

Страница 75: ...the ENT key Set a boundary point position in percentage Use the key to select a numeral and use the key to shift a digit Press the ENT key Set a boundary point position to 0500 0 Use the key to select...

Страница 76: ...icate and press the ENT key Use the key to select the channel you want to set and press the ENT key Use the key to switch from to and press the ENT key Press the ENT key Setting is completed To return...

Страница 77: ...ift a digit Press the ENT key Set the 2nd character to end in the same manner Set as the 3rd character to end and press the ENT key Press the ENT key Setting is completed To return to the user mode ho...

Страница 78: ...manner as setting the 1st character to end Repeat setting for each channel Press the ENT key Setting is completed To return to the user mode hold down the MENU key for 3 seconds or more ENT MENU ESC 3...

Страница 79: ...g Press the ENT key After reset enter the engineering mode The following describes how to set the following items The following shows a display map of the Enginneering mode Use the key to operate 7 2...

Страница 80: ...Comment 3 print Asynchronous Manual print Asynchronous Date and time print Asynchronous Setting DI No Function Setting Recording start stop INT EXT selection Tag channel print selection Alarm print O...

Страница 81: ...e Logging Print Sync ASync Print gap correction Record Start End print INT Ch OFF ON 6H 00 00 Sync print OFF OFF Use the RUN key COM communication Protocol Address Baud rate Data length Parity bit Sto...

Страница 82: ...ey to indicate and press the ENT key 7 2 2 Burnout ON OFF Setting Burnout High to deflect the recording over the span for each channel is possible Example When setting Burnout High for Channel 2 Enter...

Страница 83: ...e ENT key 7 2 4 Reference Junction Compensation The following describes how to set the method for compensating an electromotive force generated between a thermocouple wire or compensation lead wire an...

Страница 84: ...value or the key to shift a digit Once the numerical value is decided press the ENT key Setting is completed Returned to the channel setting Example When Channel 1 input compensates Channel 2 referenc...

Страница 85: ...Sets a recording start stop trigger select tag channel print or set alarm print and logging print 1 Recording start stop trigger setting Set whether the trigger of recording start stop should be by th...

Страница 86: ...n 3 Alarm print ON OFF Setting the alarm print ON OFF When ON1 prints the alarm occurring and the alarm recovering When ON2 prints only the alarm occurring Example When sets both the alarm occurring a...

Страница 87: ...to select and press the ENT key Use the key to display and press the ENT key Use the key to alter a numeral and key to shift a digit Then set the time and press the ENT key Use the key to display and...

Страница 88: ...ed to the print gap correction Setting screen 6 Record Start End Print OFF SYNC ASYNC Set Record Start End print to OFF Synchronous print Asynchronous print Example When setting Record Start End print...

Страница 89: ...o indicate and press the ENT key Select from Use the key to indicate and press the ENT key Select from 1200 2400 4800 9600 19200 38400 Use the key to indicate and press the ENT key Select from 7 bit 8...

Страница 90: ...ey to indicate and press the ENT key Use the key to select and press the ENT key Use the key to indicate and press the ENT key Setting is completed Returned to the DI3 Function Setting screen ENT PRIN...

Страница 91: ...point of 1 channel pen recording Enter the engineering mode See the key operation in page 78 Use the key to indicate and press the ENT key Use the key to select and press the ENT key Use the key to se...

Страница 92: ...T key Display indicates the set counter value Setting is completed Returned to the Span Point Calibration screen 7 2 12 Data Calibration Calibrate the voltage resistance temperature detector and refer...

Страница 93: ...ess the ENT key Enter 35 mV After 10 seconds make sure that the ALM lamp unilluminated and press the ENT key Enter 55 mV After 10 seconds make sure that the ALM lamp unilluminated and press the ENT ke...

Страница 94: ...h the key and if incorrect select respectively Then press the ENT key Display Operation Keys Description ENT ENT PRINT PRINT ENT ENT PRINT ENT ENT ENT ENT PRINT Note During calibration ALM lamp indica...

Страница 95: ...on in page 99 If the calibration is correct select with the key and if incorrect select respectively Then press the ENT key 7 3 Terminating the Engineering Mode Example The following describes how to...

Страница 96: ...peed etc Controlling of the recorder Partially Selection to print recording start stop and changing display etc For the functions to see the separate Communication Command Reference Manual charged 8 1...

Страница 97: ...nsmitting function does not reply The data format error status can be confirmed by the ESC S command described later 4 Control codes ESC 1B HEX hexadecimal This code is used for an escape sequence suc...

Страница 98: ...command is received the recorder transfers the process variable data to the transmit buffer when the ESC T command is received 2 Updating the Data ESC Once this command is received the recorder transf...

Страница 99: ...ata S Skip All the data are spaces Data information 2 1 byte E Final data Space Other data to Alarm information 1 byte each 4 bytes in total H Upper limit alarm L Lower limit alarm Space Alarm OFF or...

Страница 100: ...EVEL3L LEVEL H Upper limit alarm bit at the alarm level x LEVEL L Lower limit alarm bit at the alarm level x S Process variable data to 2 byte hexadecimal data The higher and lower bytes can be revers...

Страница 101: ...nel to Output end channel to When the number of channels is one specify S_CH E_CH Example LF 01 02 CR LF The unit and decimal point of 01 to 02 channel output 4 Data Format Data information 1 1 byte N...

Страница 102: ...nd Receiving this command outputs the status for the command received so far 2 Status Output Table 8 1 Status output list Status factor Status output Status factor A D END Syntax error Paper empty The...

Страница 103: ...esignation BO BINARY Mode only Close Command Open the Link Prepares For the Process Variable Data Output Prepares the Update Data Sends the Update Data Determines the Process Variable Data Output Orde...

Страница 104: ...tinent function code if an error occurs during command transmission an error response is returned in the fixed format described in this section Example of slave response function code 06H command erro...

Страница 105: ...s Component Data length Data Slave address 1 Function code 1 04H Data Relative start address high order 1 00H Relative start address low order 1 32H Read data count high order 1 00H Read data count lo...

Страница 106: ...version 6 16 6 30015 E Software version 7 16 7 30016 F Software version 8 16 8 30017 10 Software version 9 16 9 30018 11 Software version 10 16 10 30019 12 Software version 11 16 11 30020 13 Software...

Страница 107: ...printing 30062 3D Reserve 30100 63 Reserve 30101 64 Channel status CH01 00bit Alarm 1 1 ON 0 OFF 01bit Alarm 2 1 ON 0 OFF 02bit Alarm 3 1 ON 0 OFF 03bit Alarm 4 1 ON 0 OFF 30102 65 CH02 30103 66 Rese...

Страница 108: ...9 Unit 4 4 30139 8A Reserve 8 3 5 Reading from or Writing to the Holding Register Area The holding register area is a read write area Parameter settings and the start and stop command of the recording...

Страница 109: ...en recording deflected Are recording indication done properly Any big error in indicated values or pen recording position Any improper pen recording Any blurred pen recording or printout Proper printo...

Страница 110: ...he reference With the apparatus left in the working conditions make adjustment according to the instructions in 7 2 11 Point calibration See Page 90 9 5 Calibration Data Calibration Calibrate a proces...

Страница 111: ...Detector Conduct calibration according to the instructions in 7 2 12 Data Calibration Calibration of the resistance temperature detector See Page 93 Precision Dial Resistor TRM 10C Recorder Input Ter...

Страница 112: ...part for the difference with true value 0 5 as the right RJC temperature 24 5 0 5 25 0 Conduct calibration according to the instructions in 7 2 12 Data Calibration Calibration of reference junction co...

Страница 113: ...ment or our sales representative Remedy Trouble Does not work at all Big error Recording error Pen recording deflected Nothing is printed at all Faint printing Chart paper feed error Chart paper is no...

Страница 114: ...to 111 Set properly See 7 1 1 Scaling on Page 54 Pen recording position of Zero Span properly Calibrate again See 7 2 11 on Page 90 Recording error Flow chart 3 Pen recording deflected 0 or 100 Input...

Страница 115: ...instrument or our sales representative Take remedy according to 10 2 on Page 116 Select the RUN mode See 2 2 2 on Page 12 2 2 1 Set properly See 5 3 on Page 34 Ribbon fed properly Replace the ribbon...

Страница 116: ...er fed by FEED key DI contact for remote operation OFF Change Chart speed Turn on the DI contact See 7 2 9 on Page 89 Flow chart 7 Chart paper feed improper Chart cover locked properly Chart paper set...

Страница 117: ...ading error Engineering data EEP READ3 Non volatile memory Reading error ADC Calibration data RJC ERROR The measurement of the RJC internal temperature sensor is abnormal Operation error Calibration e...

Страница 118: ...multiple errors are encountered their error numbers are displayed with automatic scrolling When an error restores an error display can be reset to the input of RUN key CAUTION While indicates an erro...

Страница 119: ...t Allowable signal source resistance 10k max in mV TC input without burnout 100 max in mV TC input with burnout 1k max in Voltage input 10 max per line in Resistance temperature detector input Normal...

Страница 120: ...onsumption 100V Max power consumption 1 pen About 20VA About 30VA 2 pen About 25VA About 35VA 11 1 5 Normal Operating Conditions Ambient temperature 0 to 50 Ambient humidity 20 to 80 RH Supply voltage...

Страница 121: ...5 Setting accuracy Setting accuracy 11 1 7Low Voltage Directive and EMC Directive Low Voltage Directive EN61010 1 EMC Directive EN61326 1 Class A At the EN61000 4 3 test condition DC voltage DC curre...

Страница 122: ...1 K E J T C N PL U L 0 5 Table 11 2 Measurement Range Type RANGE Measurement range Measurement Digital Analog Accuracy Max resolution Accuracy DC voltage DC contact Input 10 to 10mV 0 2 of rdg 3digit...

Страница 123: ...0 15 of rdg 1 328 0 to 1112 0 F 0 15 of rdg 0 7 F 328 to 148 F 0 15 of rdg 1 8 F K3 200 0 to 300 0 0 15 of rdg 0 3 200 to 100 0 15 of rdg 1 328 0 to 572 0 F 0 15 of rdg 0 5 F 328 to 148 F 0 15 of rdg...

Страница 124: ...to 572 F 67 7 F 572 to 1472 F 33 8 F PL 0 0 to 1390 0 0 15 of rdg 0 7 32 0 to 2534 0 F 0 15 of rdg 1 3 F U 200 0 to 400 0 0 15 of rdg 0 5 200 to 100 0 15 of rdg 0 7 328 0 to 752 0 F 0 15 of rdg 0 9 F...

Страница 125: ...igits Analog recording accuracy 4 1000 1000 10 digits 0 01V Example 2 When Type K is 0 0 200 0 and Scaling is 0 00 200 00 Range accuracy rating 0 15 200 0 0 3 Input 200 0 6 6 digits Scaling factor 10...

Страница 126: ...play value Normally the accuracy rating is less than 1 digit and you cannot read by significant digit 4 Difference Sum and Average The accuracy rating of Difference Sum and Average is as following for...

Страница 127: ...Wire dot one color ink ribbon Recording Width 100 Step Response Time 1 0 s max 95 response to IEC 1143 Chart Paper Length 16m Width 114 Folding width 40 The length of the Clean Chart is 12 m Chart Sp...

Страница 128: ...t Alarm function Logging print ON OFF Reference time Interval Logging print Sync ASync Print gap correction ON OFF Alarm hysteresis Burn out DI function Option Manual print Prints Measurement result b...

Страница 129: ...t Alarm Print Prints Occurrence time Occurrence channel Setting No and Alarm type in purple at occurring an alarm Alarm Recovery Print Prints Recovery time Recovery channel Setting No and Alarm type i...

Отзывы: