5. Steam Trap Diagnosis 

5.1 Receive Trap Data 

The Trap Data is downloaded to the TM6 by operating of TrapManager-Pro. After 
downloading is successful, TM6 displays the Location ID and trap model. 

5.2 Take Measurement 

1)  Measurements should be taken from the inlet side 

of the trap. If the measurement point is at the 
outlet, measurement accuracy cannot be assured. 

2)  Press the probe against the measurement point. 

Measurements begin automatically once the probe 
is applied to the measurement point.   

3)  It takes 15 seconds after placing the probe against 

the measurement surface for the measurement to be complete. Be sure to hold 
the probe perpendicular and steady for this entire period. 

5.3 Automatic Judgment 

1)  After 15 sec measurement, TM6 confirms some parameters. 

You can change the parameters if needed. 

2)  TM6 displays a test result. 

You can modify the test result if needed. 

5.4 Upload Inspected Data 

1)  After the test result is fixed, the inspected data is transmitted to TrapManager-

Pro automatically. 

2)  After transmitting is successful, the display of TM6 returns to the normal. 

Trap Inlet 

 

Содержание TM6

Страница 1: ...RN070405A_TM6Manual_01E doc TrapMan TM6 Copyright 2007 by TLV CO LTD All rights reserved ISO 9001 ISO 14001 Manufacturer Kakogawa Japan is approved by LRQA LTD to ISO 9001 14001 ...

Страница 2: ... Diagnosis 7 5 Steam Trap Diagnosis 8 6 Settings 9 7 Specifications 11 8 Regulatory Compliance 12 1 TM6 Standard Set 1 Instruction Manual this manual 2 TrapMan TM6 3 Soft Case 4 Earphone 5 Batteries 2 alkali AAA LR03 or Ni MH rechargeable batteries 6 Cap 7 Carrying Strap Earphones Batteries Carrying Strap Cap TrapMan TM6 Soft Case ...

Страница 3: ...cepts no responsibility for any accidents or damage occurring as a result of failure to observe these precautions Symbols Indicates a DANGER WARNING or CAUTION item DANGER Indicates an urgent situation which poses a threat of death or serious injury WARNING Indicates that there is a potential threat of death or serious injury CAUTION Indicates that there is a possibility of injury or equipment pro...

Страница 4: ...e power ON check to see whether a loud sound is being output before putting on the earphones PROHIBITED Do not subject the unit to strong impact and do not throw it Such handling could result in leakage of the battery fluid excessive heat generation or injury PROHIBITED Do not place components in microwave ovens or high pressure vessels and do not place components in the vicinity of electromagneti...

Страница 5: ...tion could result in leakage of the battery fluid excessive heat generation rupture or ignition WARNING MANDATORY Cease use of the batteries immediately if the unit exhibits abnormal operation If irregularities such as fluid leakage an unusual smell unusual heat generation discoloration or deformation are noticed cease use of the unit immediately If use is continued under such conditions excessive...

Страница 6: ...RN070405A_TM6Manual_01E doc 5 3 Components Cap Ultrasonic sensor Temperature sensor LCD Key Pad LED Indicator Name Plate Rock lever Battery Pack Front view Back view Earphone Jack ...

Страница 7: ...ck clockwise to lock Locked Position Unlocked Position Battery Pack CAUTION When not intending to use for an extended period of time remove the batteries from the unit Failure to do so could result in battery fluid leakage 4 2 Turning the Power ON OFF 1 Press ENT to turn TM6 on The initial screen will appear on the display after 2 seconds 2 Press and hold the ENT key for 2 seconds to turn TM6 off ...

Страница 8: ...pManager Pro using wireless communication Before you inspect the trap you have to connect with TrapManager Pro and download the trap data to TM6 1 The wireless communication is connected by operation of TrapManager Pro 2 When TM6 and TrapManager Pro are connected the icon is shown on the TM6 display ...

Страница 9: ... automatically once the probe is applied to the measurement point 3 It takes 15 seconds after placing the probe against the measurement surface for the measurement to be complete Be sure to hold the probe perpendicular and steady for this entire period 5 3 Automatic Judgment 1 After 15 sec measurement TM6 confirms some parameters You can change the parameters if needed 2 TM6 displays a test result...

Страница 10: ...he communication with PC To communicate with TrapManager Pro you must select RF Com 6 3 Auto Off This feature automatically turns the TM6 power OFF after the prescribed length of time has elapsed in which the TM6 has not been operated a Press the and buttons to raise or lower the power OFF time delay from 0 to 15 minutes b If the time delay is set to 0 minutes the automatic power OFF feature is di...

Страница 11: ...creases the contrast level more rapidly c Once contrast is as desired press ENT once to proceed to the next setting 6 6 Language Select the language desired for displaying text 6 7 Date Time Adjust to the current date and time a Pressing once increases decreases the year month date hour and minute b Pressing ENT once to proceed to the next setting 6 8 Measurement Count This information is used for...

Страница 12: ...er Source 2 AAA LR03 batteries Suitable Battery Types Manganese batteries Alkali Batteries Ni Cd rechargeable Ni MH rechargeable RF Communication Frequency Range 2 4 2 4835 GHz The transmission scheme Frequency Hopping Spread Spectrum FHSS TX Power 4dBm Max Earphone Output Output Connection 3 5 stereo output Output Sound Tone beat Operating Temp Conditions ambient In Use 0 40 C 32 104 F In Storage...

Страница 13: ... designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harm...

Отзывы: