TKA SCIENTIFIC INSTRUMENTS

 

 
 
 
 
 
 

 

 
 

 
 
 

UV-Radiometer 

ТКА-UV  

 
 

O p e r a t i o n   M a n u a l  

 
 
 
 
 
 
 
 
 
 
 
 

 

St. Petersburg 

2008 

 
 
 
 
 
 

Generated by Foxit PDF Creator © Foxit Software

http://www.foxitsoftware.com   For evaluation only.

Содержание UV-Radiometer

Страница 1: ...TKA SCIENTIFIC INSTRUMENTS UV Radiometer UV Operation Manual St Petersburg 2008 Generated by Foxit PDF Creator Foxit Software http www foxitsoftware com For evaluation only...

Страница 2: ...are the property of the company TKA SCIENTIFIC INSTRUMENTS and protected by copyright Any full or partial reproduction of this manual s contents is prohibited This publication does not grant any righ...

Страница 3: ...stics Wight kg not above with power module 2 0 Operation parameters Ambient temperature normal operation conditions operation temperature range 20 5 0 40 Relative air humidity at ambient temperature25...

Страница 4: ...data on the irradiation levels Eee PC 701 is a 7 inch laptop which is a classic notebook with a lockless cover The laptop has no hard disk its function is performed by a 4 Gb flash memory The laptop...

Страница 5: ...t an employee of TKA SCIENTIFIC INSTRU MENTS Photometer head Laptop ASUS Eee PC 701 with power unit Cable USB2 0 connecting USB A B Documentation and registration materials Installation of the device...

Страница 6: ...of the computer For instance for the device s factory number 14003 the computer s name must be TKA 14003 Note To learn the name of the computer at which you are working you should perform the follow...

Страница 7: ...lity Cleaning the device If the device is not in use store it in a dry and clean place Clean the device once a week or when necessary Carefully wipe with a cloth without lint the eye on the photometer...

Страница 8: ...dditional adjustment by zones The buttons are not available if the device is not switched up or not found through any reason Program operation Important For the calibration it is necessary to braid in...

Страница 9: ...n the selection of the step therefore when the order is infringed failures in operation are possible The selection of the calibration step is effected through pressing the respective button Dark curre...

Страница 10: ...rement by pressing the button Measure in current range in the current window of the program When the measurement is completed the program will offer to make a measurement on the next range There are 9...

Страница 11: ...ing the range lacing ratios The ratios are calculated following the same scheme as in steps 1 and 2 Place the photometer head in parallel to the plane of the item which is being measured See to it tha...

Страница 12: ...on the bench and to make two measurements on adjoining ranges In case it is impossible to increase the signal to the ac ceptable value press the button Low Level and the remaining ratios will be calc...

Страница 13: ...evice is installed on the bench so that on the zero range the maximum signal might be obtained in the area of 3 900 4 000 units The position on the bench is fixed the photohead re moved Then in the fi...

Страница 14: ...ld of the respective UV zone the measured value is entered and calculation made Above the entry field it is necessary to place a flag for confirming the entry of changes When the flag is removed the c...

Страница 15: ...n window of the program contains the signal spectrum and energy characteristics Output of energy characteristics only can be adjusted see item Setup Generated by Foxit PDF Creator Foxit Software http...

Страница 16: ...of either display of irradiance only turned on or the mode of simultaneous display of irradiance and signal spectrum turned off In the mode of simultaneous display the irradiance output can be contro...

Страница 17: ..._YYYYMMDD_HHMMSS The obtained file can be downloaded into any word processor or electronic table for further processing Storage and transportation rules The device must be stored in a dry heated room...

Страница 18: ...cal support services the manufacturer has the right to require from you the information relating to the technical parameters of your equipment The manufacturer has the right to use the above informati...

Страница 19: ...he Software TKA SCIENTIFIC INSTRUMENTS designated representatives v publish the Software or any License Num ber or use the same other than for the purposes described in Section 1 a of this Agreement v...

Страница 20: ...sole and exclusive remedy for breach of this Warranty is replacement of the defective media or Software or at TKA SCIENTIFIC INSTRUMENTS option return of the Soft ware for a full refund In order to ex...

Страница 21: ...ter hereof d Amendments Waiver This Agreement may be modified or changed in whole or in part in any manner in writing duly signed by both parties hereto or by a further electronic agreement presented...

Страница 22: ...number on the device is modified erased removed or dam aged The warranty gives no right for recovery of indirect damage resulted from the failure of the de vices under warranty When the device fails...

Страница 23: ...St Petersburg ul Kirillovskaya 14 for correspondence P O box 234 phone 812 710 74 77 phone fax 812 274 74 43 e mail info tka spb ru http www tka spb ru Generated by Foxit PDF Creator Foxit Software h...