TE-5000D TSP
7
Operations Manual
Electrical Installation
Installation must be carried out by specialized personal only, and must adhere to all
local safety rules. This unit can be used for different power supply versions; before
connecting this unit to the power line, always check if the voltage shown on the
serial number tag corresponds to the one on your power supply. This product does
use grounded plugs and wires. Grounding provides the path of least resistance for
electrical currents, thereby reducing the risk of electric shock to users. This system
is equipped with electrical cords with internal ground wires and a grounding plug.
The plug must be plugged into a matching outlet that is properly installed and
grounded in accordance with all local codes and ordinances. Do not modify the plug
provided. If plug will not fit outlet, have the proper corresponding outlet installed by
a professional, qualified electrician.
Do Not Abuse Cords
In the event that any electrical component of this system needs to be transported,
DO NOT
carry the unit by its power cord or unplug the unit by yanking the cord
from the outlet.
Pull the plugs, not the cords
, to reduce risk of damage to unit.
Keep all cords away from heat, oil, sharp objects, and moving parts.
Extension Cords
It is always advisable to use the shortest extension cord possible. Grounded units
require a three-wire extension cord. As the distance from the supply outlet
increases, you must use a heavier gauge extension cord. Using extension cords with
inadequately sized wires results in serious changes in voltage, resulting in a loss of
power and possible damage to equipment. It is recommended to only use 10-gauge
extension cords for this product. Never use cords that exceed one hundred feet.
Outdoor extension cords must be marked with the suffix “W-A” (or “W” in Canada)to
indicate that it is suitable for outdoor usage. Always ensure that extension cords are
properly wired and in good electrical condition. Always replace damaged extension
cords immediately, or seek repair from qualified electricians before further use.
Remember to protect extension cords from sharp objects, excessive heat, and damp
or wet conditions.
Содержание TE-5000D
Страница 2: ...TE 5000D TSP 2 Operations Manual TE 5000 D TSP ...
Страница 37: ...TE 5000D TSP 37 Operations Manual Assembly Drawings TE 5005 Blower Motor Assembly ...
Страница 38: ...TE 5000D TSP 38 Operations Manual TE 5003 Filter Holder Assembly ...
Страница 39: ...TE 5000D TSP 39 Operations Manual Calibration Worksheet ...
Страница 40: ...TE 5000D TSP 40 Operations Manual Calibrator Certificate ...