![Tintometer Lovibond SensoDirect 150 Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/tintometer/lovibond-sensodirect-150/lovibond-sensodirect-150_instruction-manual_1118457022.webp)
22
SensoDirect 150
pH Calibration
11 to 3 point calibration using
pH 7 / pH 4 / pH 10 buffer solutions
Three point calibration ensures best linearity and accuracy.
Measurement Range
Resolution Accuracy*
pH
0 bis 14 pH
0,01 pH
± (0,02 pH + 2 digits)
mV
- 1.999 bis 1.999
mV
1 mV
± (0,5 % pH + 2 digits)
* pH accuracy is based on calibrated meter only.
•
Conductivity
Conductivity probe
Carbon rod electrode for long life
Function
• Conduction (µS, mS)
• Total Dissolved Solids (ppm)
• Temperature (°C, °F)
Temperature Compensation
Automatic from 0 to 60°C (32 – 140 °F),
with temperature compensation factor variable
between 0 to 5 % per °C
Probe Operating Temperature
0 to 60 °C (32 to 140°F)
Probe Dimension
Round, 22 mm diameter x 120 mm length
Probe Weight
approx. 65 g
Range
Measurement
Resolution
Accuracy
200 µS
0 bis 200,0 µS / cm
0,1 µS / cm
± (2 % F.S.
+ 1 digit)
2 mS
0.2 bis 2,000 mS / cm
0,001 mS / cm
20 mS
2 bis 20,00 mS / cm
0,01 mS / cm
200 mS
20 bis 200,0 mS / cm
0,1 mS / cm
F.S. = Full Scale
•
TDS (Total Dissolved Solids)
Range
Measurement
Resolution Accuracy
200 ppm
0 bis 132 ppm
0,1 ppm
± (2 % F.S. + 1 Stelle)
2.000 ppm
132 bis 1.320 ppm
1 ppm
20.000 ppm
1.320 bis 13.200 ppm
10 ppm
200.000 ppm
13.200 bis 132.000 ppm
100 ppm
F.S. = Full Scale