15
sage is shown and the original calibration value is re-
tained.
• User calibration
cELL
• Manufacturing calibration
CELL
9.1
Restoring the factory calibration setting
You can set the conductivity meter back to the factory
setting by simultaneously pressing the [On/Off] and
[Cal] keys when you switch the system on.
10. Result memory
A maximum of 20 results with the corresponding tem-
peratures can be stored.
10.1 Storing results
You can store the current result in the result memory
by pressing the [Store] key.
[Stor] and the number of the memory location appear
briefly in the display.
10.2 Displaying stored results
You can call the stored results up in the display by
pressing the [Rcl] (=Recall) key.
The first time you press the [Rcl] key, the number of
the memory location and the corresponding value are
displayed.
The second time you press the key, the temperature
for this result is displayed. You can use the
[
▲
]
[
▼
]
keys
to switch between the stored results.
On
Off
Cal
Store
+
Rcl