MODEL 58-FMC
FLOW METER CALIBRATOR

INSTRUCTION MANUAL

ISSUE 1217

www.tinsley.co.uk

2

DIRECTIONS FOR USING THE FLOW CALIBRATOR 58-FMC

Connect the Quick Connect to the high pressure port of DP Transmitter under calibration.

•  Leave the low pressure port of the transmitter open to atmosphere.

•  Connect the output from the transmitter through required power supply to an accurate digital multimeter.

•  Divide the range of the transmitter into five equal divisions, example 6000 mm WC =0, 1500 mmWc -25%, 

3000 mmWC – 50%, 4500 mmWC – 75% and 6000 mmWC – 100%.

•  Check the Zero output of the transmitter without giving any pressure to the HP and LP side.

•  Give 6000 mmWC (full span 100%) to the HP side, check output correct span if required.

Check zero for any shift and correct if necessary.

•  Close the return valve of the squeeze bulb and pump pressure using the squeeze bulb to 25% (eg. 1500 mmWC) 

check the output.

•  Go on in steps of 25% to check intermittent calibration.

•  Once again check zero and span for any shift or error.

•  Tabulate readings for future reference.