OPERATIONAL CHECK
Prior to use, the Model AP/W Holiday Detector should be given an operational check to
ensure it is working properly. The Operational Check ensures that the audible and visual
alerts are working properly and the instrument has voltage output. The Operational Check is
not a calibration check and should not replace the recommendation of annual service.
The instrument should be completely assembled and ready for use in the application in
which it is required.
Operational Check
•
Ensure the instrument and all components are assembled.
•
Visually inspect each cable and connection.
•
Squeeze the safety handle on the PowerPak.
•
Press and Release the Power button on the top of the PowerPak.
•
Note the Power LED on the Powerpak near the Power button. It should be on with a
steady glow. A slow blinking of the Power LED indicates a Low Battery condition.
•
An audible alert should be noted when the instrument turns On.
•
A visual alert on the top of the PowerPak should be seen when the instrument turns
On.
On the panel of the AP/W instrument:
•
Note the LCD Voltage display does not fluctuate and change significantly. A small
letter “b” on the display indicates a Low Battery condition.
•
Note the Power LED near the Power button remains steady and not flashing. A slow
blinking indicates a Low Battery condition.
•
Touch the electrode to the bare ground wire. Note the audible and visual alerts.
Repeat this step several times.
NOTE: Spark may or may not be seen and heard. When using lower voltage
outputs and on bright sunny days, the spark may not be seen or heard.
•
Release the safety handle to turn the instrument Off.
The Model AP/W is now ready to use.