Timex 80222 Скачать руководство пользователя страница 4

4

Note: Printer may be purchased by calling 1-800-422-0360 or on-line at www.timex-
healthcare.com.

How to print the last stored measurement. (Single data print)

a) Select correct printer objective first before you press the PRINTER button.

b) Connect the printer with the monitor and switch the printer on (for details please
refer to the manual of the printer).

c) Press the PRINTER button of the monitor once. The display will show "Pr1" and the
printer will print the last stored measurement for User 1 or 2 based on your selection.

How to print all measurements stored in memory and a data chart, (data
print out with chart).

a) Select correct printer objective first before you press the PRINTER button.

b) Connect the printer with the monitor and switch the printer on (for details please
refer to the manual of the printer).

c) Press the PRINTER button for at least 3 seconds. The display will show "Pr1" fol-
lowed by "Pr2".

d) Release the button, and the printer will print all stored measurements with a corre-
sponding chart.

Stop printing

If you would like to stop printing, press the PRINTER button during the printing
process.

NOTE: Further details regarding the printer can be found in the printer instruction
manual.

ERROR MESSAGES/MALFUNCTIONS

If an error occurs during a measurement, the measurement is discontinued and a cor-
responding error code is displayed.

ERROR CHART

Other possible malfunctions and possible solutions: If problems occur when using the
device, the following points should be checked and if necessary, the corresponding
measures are to be taken:

Malfunctions Remedy Chart

If an error occurs during a measurement, the measurements are 
discontinued and a corresponding error code is displayed. (example: error no. 1)

Error No.

Possible cause(s)

Err 1

The systolic pressure was determined but afterwards the cuff 
pressure fell below 20 mmHg. This situation can occur, for 
example, if the tube becomes unplugged after the systolic 
blood pressure has been measured. Further possible cause:
No pulse has been detected

Err 2

Unnatural pressure impulses impair the result of the 
measurement. Cause: The arm was moved during the 
measurement.

Err 3

Inflation of the cuff takes too long. The cuff is not correctly 
seated or the hose connection is not sealed.

Err 5

The measured readings indicated an unacceptable difference 
between systolic and diastolic pressures. Take another 
reading taking care to follow directions carefully. Contact 
you doctor if you continue to get unusual readings.

Err 6

Single data differ too much during the AAT Mode even after 
4 cycles. No average result can be displayed.

Other possible malfunctions and their elimination

If problems occur when using the device, the following points should be checked and,
if necessary, the corresponding measures are to be taken:

Malfunction

The display remains empty when the 
instrument is switched on although 
the batteries are in place.

Remedy

1. Check whether the batteries are
installed with the correct polarity and 
correct if necessary. 2. If the display is
unusual, remove the batteries and 
then exchange them for new.

Malfunction

The pressure does not rise although 
the pump is running

The device frequently fails to measure 
the blood pressure valves, or the valves 
measured are too low (too high)

Every measurement results in different 
values, although the device functions 
normally and normal values are displayed.

The blood pressure values measured 
differ from those measured by the doctor.

Further information

The level of blood pressure is subject to fluctuations even in healthy people. It is
important to compare measurements taken under the same conditions and at the
same time of day. (Quiet conditions) The device has been subjected to strict clinical
tests by experienced Cardiac Specialists. This device is manufactured to meet
European standards under the supervision of the Technical Monitoring Association.
(RWTüV-Essen).If you have any questions regarding the use of this blood pressure
monitor, please ask your pharmacist or contact the TIMEX customer service represen-
tative, 1-800-422-0360. The TIMEX customer service team will be happy to help you.
Never attempt to repair the instrument yourself! Any unauthorized opening of the
instrument invalidates all warranty claims!

CARE AND MAINTENANCE

a) Do not expose the device to extreme temperatures, humidity, dust or direct 
sunlight.

b) The cuff contains a sensitive air-tight bubble. Handle the cuff carefully and avoid all
types of straining through twisting or buckling.

c) Handle the tube carefully. Avoid stress from pulling, pinching and sharp edges.

d) Clean the device with a soft, dry cloth. Do not use gasoline, thinners or similar 
solvents. Carefully remove spots on the cuff with a damp cloth and soapsuds. The cuff
should not be washed in a washing machines or dishwasher!

e) Do not drop the instrument or treat it roughly in any way. Avoid strong vibrations.

f) Never open the device! This can negatively impact the calibration!

WARRANTY

Your TIMEX™ Professional Series Upper Arm Blood Pressure Monitor is a durable and
precise medical device manufactured under license by MEDport, LLC. MEDport, LLC.
takes pride in the quality and workmanship crafted into every product made.
Accordingly, MEDport, LLC. will repair or replace at our option any product that is
defective in material or workmanship for three years. This warranty does not cover
the batteries or damage caused by accident, negligence or improper use or care.
Please take a moment to mail in the enclosed warranty card now. It will expedite any
claims you may have against this warranty in the future.

Product should be returned to MEDport, LLC., postage prepaid with proof of purchase
and an explanation of the problem. If you follow the instructions for proper use and
care, the TIMEX™ Professional Series Upper Arm Blood Pressure Monitor should last
for years. However, if for any reason you are not satisfied, call our toll-free customer
service line: 1-800-422-0360. Our hours of operation are Monday – Friday 8:00 a.m.
– 8:30 p.m. EST.

The warranty is only valid upon presentation of the warranty card filled out by the
consumer and on file.

REFERENCE TO STANDARDS

Device standard:

Device corresponds to the requirements of the 
European standard for non-invasive blood 
pressure monitor

EN1060-1 / 12:95
EN1060-3 / 09:97

Remedy

1. Check the connection of the cuff tube
and connect properly if necessary.

1. Fit the cuff correctly on the arm.
2. Before starting the measurement make
sure that the cuff is not fitted too tightly
and that there is no rolled-up sleeve 
exerting pressure on the arm above the
measuring position. Take off articles of
clothing if necessary.

Please read the following information 
and the points listed under “Common
sources of error” on page 12. Repeat 
the measurement.

Record the daily development of the 
values and consult your doctor about
them.

Отзывы: