4 Preparation
4-8
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
4.5 Filling of the reservoir during start up
If the level in the reservoir is below 75 mL, topping up will be necessary even for the system
test. Wait for a filling prompt (message on the screen) and follow the instructions.
Comment:
You can find a list of the available bottle adapters in Chapter 9 or at
https://tim-gmbh.de/en/bottle-adapter/
Comment:
You can find further information on the bottle adapter in the manufacturer’s Instructions for
use.
CAUTION
Hazard to persons if volatile anaesthetic accumulates in the
environment around the patient.
Observe the Instructions for use of the bottle adapter.
Avoid spilling liquid volatile anaesthetic.
The fill valve below the fill port flap must never be pressed down
manually.
IMPORTANT Equipment damage caused by using bottle adapters not compatible
with the MIRUS Controller filling system.
Only use compatible bottle adapters.
This means ensuring that the colour coding of the bottle adapter,
the fill port collar, and the bottle matches that of the volatile
anaesthetic being filled.
1. Attach the bottle adapter to the bottle.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...