LO) will then be displayed with or without buzzer
.
If a relay board is installed, the comparison results are also sent through the
relay output at the back of this instrument.
Follow the below steps to trigger check weighing mode: -
a. During normal operation (of a function), press
[CHECK]
.
b. The current HI limit is displayed with the
Hi symbol
on, press
[TARE]
to confirm, or
c. Enter a new HI limit through the numeric keys and then press
[TARE]
,
d. Display current LO limit with the
Lo symbol
on, press
[TARE]
to
confirm, or
e. Enter a new HI limit through the numeric keys and then press
[TARE]
f.
Check weighing function is now enabled. The check is result is shown
by one of the HI/OK/LO symbols,
g. The same result will be sent to the relay output (if ordered) together
with the buzzer signal (if ordered).
NOTE: -
1. For normal comparison, set both HI and LO limits,
2. To check only if result is lower or equal to LO (result
≤
LO?), set HI limit
= 0,
3. To check only if result is higher or equal to HI (result
≥
HI), set LO limit
= 0,
4. To check if result is equal to a specified value, set both HI limit and LO
limit = the specified value
To quit check weighing mode, set both HI and LO limits to zero.
6.11. Quick setting
Press [SET]
to get quick setting parameter ,Press
[TARE]
to enter and
[CNT] and
[UNIT] to shift : PARA,CHECK,ANIMAL,RS232, and
RECHARGE
: To check the currency when charging(mA)
Count :
Display the calibration count value, and the parameter set count
value, (press
[ZERO]or [TARE] for 3secs to exit
)
17