Operating instructions

DC current measurements

1. Set the slide switch to 40 A DC or 400 A DC
2. Connect the negative output of the Current Clamp TP-CC400

(black) to the negative side of the input of the measuring in-
strument. Connect the positive output of the Current Clamp
TP-CC400 (red) to the positive side of the input of the mea-
suring instrument.

3. Switch the measuring instrument on and set it to 200 mV

DC, 400 mV DC or 2 V DC full scale input range.

4. With firmly closed jaw, press the zero button until the in-

struments reads zero.

5. Open the jaw and fully enclose the current-carrying conduc-

tor(s) and close the jaw again, no air gap is allowed between
the two halves. The output voltage is positive when the cur-
rent flows from the underside to the upside of the clamp.

6. Read the measured voltage value and convert it to the corre-

sponding current value:
40 A range: 10 mV equals 1 A
400 A range: 10 mV equals 10 A.

AC current measurements

1. Set the slide switch to 40 A AC or 400 A AC
2. Connect the negative output of the Current Clamp TP-CC400

(black) to the negative side of the input of the measuring in-
strument. Connect the positive output of the Current Clamp
TP-CC400 (red) to the positive side of the input of the mea-
suring instrument.

3. Switch the measuring instrument on and set it to 200 mV

AC, 400 mV AC or 2 V AC full scale input range.

4. Open the jaw and fully enclose the current-carrying conduc-

tor(s) and close the jaw again, no air gap is allowed between
the two halves.

5. Read the measured voltage value and convert it to the corre-

sponding current value:
40 A range: 10 mV equals 1 A
400 A range: 10 mV equals 10 A.

Current Clamp TP-CC400

5

Содержание TP-CC400

Страница 1: ...Current Clamp TP CC400 User s manual Revision 1 2...

Страница 2: ...nductors with voltages equal to or exceeding 300 V DC or 240 Vrms AC To avoid physical injury measurements on bare conductors or conductors with damaged insulation are not allowed Check the Current Cl...

Страница 3: ...C400 for which this declaration is valid is in compliance with EN 55011 2009 A1 2010 EN 61000 6 1 2007 EN 55022 2006 A1 2007 EN 61000 6 3 2007 according the conditions of the EMC standard 2004 108 EC...

Страница 4: ...ton This buton sets the current readout to zero This function is also used when measuring DC current to remove an offset value caused by residual magnetism in the core of the jaw 5 Power ON OFF led 6...

Страница 5: ...from the underside to the upside of the clamp 6 Read the measured voltage value and convert it to the corre sponding current value 40 A range 10 mV equals 1 A 400 A range 10 mV equals 10 A AC current...

Страница 6: ...test leads from the Current Clamp TP CC400 2 Remove the screw from the battery compartment at the back of the Current Clamp TP CC400 3 Remove the old battery 4 Insert a new 9 V NEDA 1604 IEC 6F22 bat...

Страница 7: ...attery Operating environment 0 C to 50 C 70 RH Storage environment 20 C to 70 C 80 RH Compliances CE Yes ROHS Yes Accuracy At 23 C 5 C 70 RH maximum DC 40 A range 0 A to 20 A 3 5 6 digits 20 A to 40 A...

Страница 8: ...8601 WL Sneek The Netherlands Tel 31 515 415 416 Fax 31 515 418 819 E mail sales tiepie nl www tiepie com This information is subject to change without notice Copyright c 2017 TiePie engineering All r...

Отзывы: