Preventative Maintenance Checklist (81000)
38
Zero-Gravity
®
Radiation Protection System Preventative Maintenance Checklist
4. Replace the leaded acrylic shield if any of the abnormalities are observed.
Contact a TIDI Products authorized representative or TIDI Products service for
repairs and / or replacement of components. Replacement and repairs may only
be completed by TIDI Products authorized personnel.
5. If additional support is needed, contact TIDI Technical Service using the provided
contact information.
Moving Parts Inspection (ZGCMRS)
Rotational / Translational Motion Parts Replacement
1. The ZGCMRS contains several components allowing for the movement
(rotational and translational motion) of other components.
2. As a result of long-term use, these components may require replacement after an
extended period of time to ensure the system functions as intended. Complete
inspection to determine if replacement of any of the components is required.
Rotational / Translational Motion Parts Inspection
1. While the system is not in use, slide the system along the usable space of the
monorail to inspect for motion issues. The system should move freely along the
monorail with minimal interruption. Other components may need to be moved in
order to test the full range of the monorail.
2. Rotate each of the components at the rotation points both clockwise and
counterclockwise. (Figure 38) The components should rotate freely with minimal
interruption.
Note that some of the rotation points have specific stop points
preventing over rotation.
3. Move the arm of the system up and down to inspect for motion issues. The arm
should move freely with minimal interruption.
4. Contact a TIDI Products authorized representative or TIDI Products service for
repairs and / or replacement of components. Replacement and repairs may only
be completed by TIDI Products authorized personnel.
5. If additional support is needed, contact TIDI Technical Service using the provided
contact information.