39
Error messages and malfunction
Problem Possible
Cause
Correction
high measurement error
contamination, ageing,
extreme temperature,
maladjustment
readjustment
display shows "ur"
pressure under range
(pressure < 10
-3
mbar)
error message "Err"
adjustment done at wrong
pressure
measurement error out of
adjustment range
displayed pressure must be
>40mbar f. atmosphere
adjustment, <4x10
-2
mbar f.
zero adjustment
exchange sensor head
error message "Err1"
defective sensor
no sensor connected
exchange sensor head
check sensor connection
Содержание VD84M
Страница 2: ...VD84M VD84MIR Kompakt Vakuummeter Compact Vacuummeter Betriebsanleitung Operating Instructions...
Страница 22: ...21 Konformit tserkl rung...
Страница 42: ...41 Declaration of Conformity...
Страница 43: ......
Страница 44: ......
Страница 45: ......