background image

CRESCENT 

CONNECTOR

LATERAL

PANEL

(2)

ANTERIOR

PANEL

ASSEMBLY

S

LEEQ

MAX

 

TLSO

38 - 44

44 - 54

32 - 38

26 - 32

80 - 95

65 - 80

95 - 110

110 - 136

4

3

2

1

1

2

3

4

Sizing Arqs

Sizing Arqs

PULL TAB

1.1

1.0

SIZING

Indications:

Sagittal/coronal/transverse plane motion restriction; increased intracavitary pressure to reduce spine 

& disc loading; mechanical or discogenic thoracic/lumbar & extremity pain; post-operative spinal 

stabilization/fusion; degenerative disc disease; post-injury to facilitate healing; spondylolysis; spinal 

stenosis; spondylolisthesis; spondylosis; laminectomy; discectomy; facet syndrome; spinal muscle 

weakness; arthritis; low back sprains and strains; prophylactic back support.

Contraindications:

• SLEEQ braces are contraindicated for women who are pregnant.

• SLEEQ braces are contraindicated for persons with circulatory, pulmonary, cardiovascular or other 

conditions where increased abdominal pressure would be of concern.

Prepare to Fit: 

Unclip straps (1) from sternal plate and remove anterior panel 

assembly (2) from wing.

2.1

3.1

3.2

3.3

3.4

2.0

ATTACH WING SHELLS & LATERAL PANELS

3.0

ADJUST & FIT

 ALIGN

 wing shell on wing with sew line Arq on shell aligned to sew line Arc on  

wing where crescent connector is attached from step 1.2 (see arrows below).

1.2

Adjust Circumference:

  Measure waist circumference and reference size chart 

above for proper alignment.

2

REMOVE

PUSH

1

SIZE

REFERENCE

RECEIVER

CLIP

WING SHELL 

(2)

IN.

CM.

Move right wing in or out and align Crescent Connector to appropriate Sizing Arq.

Repeat on left wing.

Scan To View 

Instructional Fitting 

Video

LIFT

1. ALIGN

2. TUCK

3. SECURE

WRAP

WRAP

(A)

• Position the brace with lower edge of back panel at coccyx level. 

• Position the anterior panel with lower edge just above symphysis pubis level.
• Wrap the left wing (1) and attach to Velcro on the anterior panel, then wrap 

right wing (2).

4

1

2

3

4

LIFT

SLIDE

SYMPHYSIS

PUBIS

STERNAL

NOTCH

STERNAL BAR CLIP

Note: 

When fitting in 

standing position, patient 

should hold wings at 

their sides to allow 

practitioner to place 

anterior panel centered  

at midsagittal line.

Tighten brace:

 Detach pull-tabs, pull out & away from the body then 

re-attach anywhere onto the wings. (

See section 5.0 for resetting the 

tighening mechanism)

UNDER AXILLA STRAPPING OPTION

(Also see MODIFICATIONS section 4.3)

COCCYX

FOLD

(A)

1

2

• Attach strap clips into sternal plate and adjust straps so that strap padding 

is centered on the anatomy.

• The strap ends should be tucked into the anterior strap clips and posterior 

strap loops. Strap ends can also be cut and heat sealed.

Sternal Bar:

 Open sternal bar clip to telescope sternal bar to desired 

height, then close clip to lock in place.

– Standard practice is to have sternal plate 1 to 2” below sternal notch 

when standing and 1/2” below when seated or supine.

– If there is a gap between sternal pad and chest or there is too much 

pressure on the chest, the sternal bar can be bent to achieve proper 

positioning with bending irons or over the edge of a table.  

Note:

 Bend 

sternal bar above toothed area.

EXT Extension Panel

(Sold Separately)

Extends SLEEQ Max to 50”-60”

(127-152 cm)

• Attach wing shell Velcro to inside of wing and secure Velcro tab (A) at inner end of  

wing inside back panel.
• 

TUCK 

and attach wing shell inside back panel as shown above.

• 

SECURE 

Velcro tab (A) to inner end of wing inside back panel.

• Repeat on other side. 

Note: 

For sizes 1, 2, 3, use center Arq on wing shell, and for size 4 use posterior 

Arq near Velcro tab (A).

ATTACH LATERAL PANELS

Отзывы: