CRESCENT
CONNECTOR
LATERAL
PANEL
(2)
ANTERIOR
PANEL
ASSEMBLY
S
LEEQ
MAX
™
TLSO
38 - 44
44 - 54
32 - 38
26 - 32
80 - 95
65 - 80
95 - 110
110 - 136
4
3
2
1
1
2
3
4
Sizing Arqs
™
Sizing Arqs
™
PULL TAB
1.1
1.0
SIZING
Indications:
Sagittal/coronal/transverse plane motion restriction; increased intracavitary pressure to reduce spine
& disc loading; mechanical or discogenic thoracic/lumbar & extremity pain; post-operative spinal
stabilization/fusion; degenerative disc disease; post-injury to facilitate healing; spondylolysis; spinal
stenosis; spondylolisthesis; spondylosis; laminectomy; discectomy; facet syndrome; spinal muscle
weakness; arthritis; low back sprains and strains; prophylactic back support.
Contraindications:
• SLEEQ braces are contraindicated for women who are pregnant.
• SLEEQ braces are contraindicated for persons with circulatory, pulmonary, cardiovascular or other
conditions where increased abdominal pressure would be of concern.
Prepare to Fit:
Unclip straps (1) from sternal plate and remove anterior panel
assembly (2) from wing.
2.1
3.1
3.2
3.3
3.4
2.0
ATTACH WING SHELLS & LATERAL PANELS
3.0
ADJUST & FIT
•
ALIGN
wing shell on wing with sew line Arq on shell aligned to sew line Arc on
wing where crescent connector is attached from step 1.2 (see arrows below).
1.2
Adjust Circumference:
Measure waist circumference and reference size chart
above for proper alignment.
2
REMOVE
PUSH
1
SIZE
REFERENCE
RECEIVER
CLIP
WING SHELL
(2)
IN.
CM.
Move right wing in or out and align Crescent Connector to appropriate Sizing Arq.
Repeat on left wing.
Scan To View
Instructional Fitting
Video
LIFT
1. ALIGN
2. TUCK
3. SECURE
WRAP
WRAP
(A)
• Position the brace with lower edge of back panel at coccyx level.
• Position the anterior panel with lower edge just above symphysis pubis level.
• Wrap the left wing (1) and attach to Velcro on the anterior panel, then wrap
right wing (2).
4
1
2
3
4
LIFT
SLIDE
SYMPHYSIS
PUBIS
STERNAL
NOTCH
STERNAL BAR CLIP
Note:
When fitting in
standing position, patient
should hold wings at
their sides to allow
practitioner to place
anterior panel centered
at midsagittal line.
Tighten brace:
Detach pull-tabs, pull out & away from the body then
re-attach anywhere onto the wings. (
See section 5.0 for resetting the
tighening mechanism)
UNDER AXILLA STRAPPING OPTION
(Also see MODIFICATIONS section 4.3)
COCCYX
FOLD
(A)
1
2
• Attach strap clips into sternal plate and adjust straps so that strap padding
is centered on the anatomy.
• The strap ends should be tucked into the anterior strap clips and posterior
strap loops. Strap ends can also be cut and heat sealed.
Sternal Bar:
Open sternal bar clip to telescope sternal bar to desired
height, then close clip to lock in place.
– Standard practice is to have sternal plate 1 to 2” below sternal notch
when standing and 1/2” below when seated or supine.
– If there is a gap between sternal pad and chest or there is too much
pressure on the chest, the sternal bar can be bent to achieve proper
positioning with bending irons or over the edge of a table.
Note:
Bend
sternal bar above toothed area.
EXT Extension Panel
(Sold Separately)
Extends SLEEQ Max to 50”-60”
(127-152 cm)
• Attach wing shell Velcro to inside of wing and secure Velcro tab (A) at inner end of
wing inside back panel.
•
TUCK
and attach wing shell inside back panel as shown above.
•
SECURE
Velcro tab (A) to inner end of wing inside back panel.
• Repeat on other side.
Note:
For sizes 1, 2, 3, use center Arq on wing shell, and for size 4 use posterior
Arq near Velcro tab (A).
ATTACH LATERAL PANELS