Figure 6 APP measurement after connected
Figure 7 Instrument measurement interface after connected APP
Страница 1: ...ction against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to ...
Страница 2: ...of light source color space color index and reflectance functions which can realize massive chromaticity index testing This instrument has the following advantages 1 It can be measured after power on no need to perform white calibration every time making color measurement easier 2 Separate design of calibration whiteboard and measuring port to ensure long service life of whiteboard and accurate me...
Страница 3: ...rface front Figure 2 Schematic diagram of button interface rear Figure 3 Schematic diagram of calibration cover Display Operate Button LED Indicator Measuring Button Power Switch Type C USB Port Tether hole Measuring Aperture Calibration Cover ...
Страница 4: ...t will automatically shut down Note When not using the instrument for a long time please turn off the power 2 3 Black and white calibration In the measurement interface long press the operation button to enter the system function interface slide the operation button to the left or right position the cursor on the black and white calibration and long press the operation button to enter the calibrat...
Страница 5: ...h or Connect to the instrument first check the instrument model and SN code in the Device Information of the instrument and then in the mobile APP Select the instrument to be connected instrument model SN code in the Connected Instruments list and click to automatically connect In the color measurement interface click standard sample measurement or take standard sample to obtain the chromaticity d...
Страница 6: ...Figure 6 APP measurement after connected Figure 7 Instrument measurement interface after connected APP ...
Страница 7: ...rement interface align the measurement aperture to the standard sample and press the measurement button shortly The instrument obtains the chromaticity data of the standard sample as shown in Figure 9 The green indicator LED flashes once when the measurement is completed The status bar at the top of the instrument screen displays the current measurement conditions the icon at the bottom displays t...
Страница 8: ...can be continued after the sample is cleared Slide the operation button to the left or right to view the sample data The default DE tolerance of the instrument is 1 0 When the color difference is within the tolerance range it will prompt Qualified otherwise it will prompt Bad The instrument DE tolerance can be set via APP and SQCX Figure 10 Instrument sample measurement 2 7 Mini printer The mini p...
Страница 9: ...ve When auto save is turned on the measurement result will be automatically stored in the instrument for each measurement Otherwise the sample measurement will end and the measurement record will not be automatically saved 3 5 2 Sound When the buzzer sound is turned on a reminder sound will sound every time you operate otherwise there will be no reminder sound 3 5 3 Bluetooth For instruments equip...
Страница 10: ... 60 seconds 90 seconds normally open If normally open is selected the screen will always be on If it is set to 30 seconds the instrument will count from the last operation and the display will turn off after 30 seconds The meanings of 60 seconds and 90 seconds are the same as above 3 5 9 Factory reset Select Restore Factory the instrument will clear all measurement records and user settings and re...
Страница 11: ...ical Geometry D 8 Diffuse illuminance 8 reception SCI Conforms to standards CIE No 15 GB T 3978 Features Used for color difference quality control in plastic electronics paint and coating textile and garment printing and dyeing ceramics and other industries Small size easy to carry search for the color cards and multifunction expanded through smartphone APP Integrating Sphere Size Φ20mm Illuminant...
Страница 12: ...il result color simulation color deviation reflectance partly through mobile APP Some models don t support reflectance Measuring Time About 1 0s Repeatability Color ΔE ab within 0 05 After warm up calibration the average value of measuring the whiteboard 30 times at an interval of 5s Accuracy Pass National Metrology Some models are not guaranteed Measuring Mode Single measurement Average measureme...
Страница 13: ...ng Temperature range 0 40 0 85 RH No condensation Altitude lower than 2000m Storage Temperature Range 20 50 0 85 RH No condensation Standard Accessories Data cable manual calibration box SQCX PC Color quality management software MOBCCS APP download from official website Some models don t have SQCX software Optional Accessories USB mini printer powder test box Note Technical parameters are for refe...