200CR Back Panel 

 

Warning: Make sure power to all wires is turned off 
before proceeding with the power installation.  High 
voltage may be present on the input power wires 
and relay wires. 

 

Terminal Block TB1 
200CR models 6220 and 6222 have 2 relays 
200CR model 6224 has 4 relays 
The wiring sequence is shown in the table below. 
Relays 3 and 4 are solid state, for AC only. Refer to 
Manual Chapter 2. 

 

TB1 
Label

 

Input Power & Relay 
          Function

 

 L 
 N 

 

115V/230VAC Line 
115V/230VAC Neutral 
Earth Ground 

 NC1 
 C1 
 NO1 

Relay 1: Normally Closed 
Relay 1: Common 
Relay 1: Normally Open 

 NC2 
 C2 
 NO2 

Relay 2: Normally Closed   
Relay 2: Common 
Relay 2: Normally Open 

 C3 
 NO3 

Relay 3: Common 
Relay 3: Normally Open 

 C4 
 NO4 

Relay 4: Common 
Relay 4: Normally Open 

 

AC Power Voltage and Frequency 
To change the power voltage and frequency from factory 
settings refer to Manual Chapters 2 and 4 respectively. 

 

Output Connections 
Connections for all outputs are made to terminal block 
TB2.  The serial port can be configured as an RS-232 
port or an RS-422 port.  Analog outputs, if included, are 
powered. Do not connecto circuits that provide external 
power. 
Output Connections continued 
 

TB2 
Label

 

   

RS232 

  Function

 

     

RS422 

   Function

 

RXD- 

Receive Data 

Receive Data - 

RXD+ 

Not Used 

Receive Data + 

TXD- 

Transmit Data 

Transmit Data - 

TXD+ 

Not Used 

Transmit Data +

GND Ground* 

Not 

Used 

AO2- 
AO2+ 

Analog Output 2 - 
Analog Output 2 + 

 

AO1- 
AO1+ 

Analog Output 1 - 
Analog Output 1 + 

  

*For RS232 only.  

CAUTION: Do not connect analog output wiring 
shields to adjacent GND terminal. Connect them to 
AC-power earth ground terminal only. 

 

Sensor Connections 
Wire sensor patch cord leads as shown below.  

Warning: Miswiring patch cords may damage 
sensors.  Blue wire #7 is not used. Leave clear 
shrink tube in place over it.

 

 

 

TB3 

Label

 

    Wire 
    Color

 

   Sensor 
Connection

 

GND(6) 
SIG5(5) 
SIG4(4) 
SIG3(3) 
SIG2(2) 
SIG1(1) 

 BLACK 
 RED 
 GREEN 
 WHITE 
 CLEAR 
 WHT/BLUE 

 
Channel B Sensor 
Connections 
 

GND(6) 
SIG5(5) 
SIG4(4) 
SIG3(3) 
SIG2(2) 
SIG1(1) 

 BLACK 
 RED 
 GREEN 
 WHITE 
 CLEAR 
 WHT/BLUE 

 
Channel A Sensor 
Connections 

Mettler-Toledo Thornton, Inc. 

 

 

 

Toll-Free: 800-510-PURE 

36 Middlesex Turnpike 

 

 

 

Fax: 781-271-0214 

Bedford, MA 01730 

 

 

 

[email protected] 

(781) 

301-8600 

 

   

www.thorntoninc.com

www.thorntoninc.com

www.thorntoninc.com

www.thorntoninc.com 

 

      OM84304 

Rev. 

08/03

 