42
HA0143T Rev 12K Sept 2015
Chapter 6
6.2.2 Control Tab
Fig. 6.1 Piezo Settings panel
Feedback Loop Settings
Loop mode
- This panel can be used to select between the two distinct
operating modes: open loop and closed loop. Note that closed loop operation
is only possible if an external position reader (such as the TGS001 Strain
Gauge Reader) is used in conjunction with the piezo cube.
When the T-Cube Piezo Driver is used in conjunction with the T-Cube Strain Gauge
Reader unit (TSG001) on the T-Cube Controller Hub (TCH002), high precision closed
loop operation is possible using the complete range of feedback equipped piezo
actuators available from Thorlabs.
Caution
If the Piezo Driver is switch to closed loop mode without being connected
to a Strain Gauge Reader, the piezo driver will expect a feedback signal
and the piezo drive voltage will ramp up to the maximum limit. If the limit
is set to 75V, a protective circuit limits the voltage to 75V. If 100V or 150V
is selected, the output will ramp to 150V.