Spectrum Analyzer Controller
Chapter 5: Operation
Page 12
6679-D02
4. Remove the shorting jumper and place across (shorting) the JP3 pins to
disable the blanking circuit. The default setting will be blanking enabled.
The jumper will not be shorting the pins.
5. Replace the enclosure cover and secure with the enclosure screws.
5.2. Replacing the Fuse
DANGER!
The Thorlabs SA201 Spectrum Analyzer Controller must be powered off,
unplugged from the AC input source, and disconnected from any piezo
elements prior to replacing the fuse or removal of the cover. Failure to do so
may cause SERIOUS INJURY to the user, since high voltages exist within the
unit.
5.2.1.
Materials Needed
•
SA201 Operating Manual
–
The most recent version of this operating
manual will be available on the Thorlabs web site.
•
250
mA Type ‘T’ Slow Blow Fuse –
The 250 mA fuse is installed from the
factory. This must be installed when operating the unit at 100/115 VAC.
•
125
mA Type ’T’ Slow Blow Fuse –
The 125 mA fuse is required for 230
V operation only. Thorlabs supplies a 125 mA fuse with all of its SA201
units and must be installed when operating at 230 VAC.
•
Phillips Head Screwdriver (#2 Preferred)
–
We do not recommend using
electrically powered screwdrivers.
5.2.2.
Fuse Replacement
IMPORTANT
Disconnect the scanning head or any piezo device from the SA201 output.
Disconnect the power cord.
Do not open the unit if the power cord is connected.
1. Remove the two screws securing the enclosure cover with a Phillips head
screwdriver. The screws are located on the bottom side, rear corners of
the unit. Do not lose the screws.
2. Carefully remove the cover by sliding toward the rear of the unit.