Chapter 2: Safety

 

Page 2

 

19146-D02 

Chapter 2 

Safety 

CAUTION 

DO NOT apply TTL or square wave modulation to the MOD IN input. Due to response delays of the 

constant power control loop, damage to the integrated fiber coupled laser may occur! 

 

SHOCK WARNING 

High voltage inside. To avoid electrical shock, before powering unit, make sure that the protective 

conductor of the 3-conductor power cord is correctly connected to the protective earth contact of the 

socket outlet. Improper grounding can cause electric shock resulting in severe injury or even death. 

Do not operate without cover installed. 

 

WARNING 

This unit must not be operated in explosive environments. 

Thorlabs provides the proper power input cable for use in the United States. If using this unit anywhere else, the 
user will need to supply a properly grounded power cable to power the unit. 

Do not obstruct the air-ventilation slots in the housing! 

Make sure that the line voltage rating marked on the rear panel agrees with your local supply and that the 
appropriate fuses are installed. Changing of the mains fuse can be done by the user (see Setting the AC Line 
Voltage and Installing Fuses). 

With the exception of the mains fuses, there are no user serviceable parts in this product.  

This device can only be returned when packed into the complete original packaging, including all foam packing 
inserts. If necessary, ask for a replacement package.  

Mobile telephones, cellular phones or other radio transmitters should not to be used within the range of three 
meters of this unit since the electromagnetic field intensity may exceed the maximum allowed disturbance values 
according to EN50082-1. 

 

LASER RADIATION 

AVOID EXPOSURE TO THE BEAM 

CLASS 1M LASER PRODUCT 

1310 - 1550 nm 

<5 mW 

 

LASER RADIATION 

AVOID EXPOSURE TO THE BEAM 

CLASS 3B LASER PRODUCT 

405, 473, 520 nm 

<25 mW 

Содержание S3FC405

Страница 1: ...S3FC405 S3FC473 S3FC520 Fiber Coupled Laser Source User Guide...

Страница 2: ...2 Turning On the Source 7 5 3 Adjusting the Laser Output Power 7 5 4 Turning the Laser Off 7 5 5 Adjusting the Temperature of the Laser Diode 7 5 6 Modulating the Laser Output 8 Chapter 6 Making the...

Страница 3: ...Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off Supply...

Страница 4: ...a properly grounded power cable to power the unit Do not obstruct the air ventilation slots in the housing Make sure that the line voltage rating marked on the rear panel agrees with your local supply...

Страница 5: ...ot temperature controlled version S3FC series comes equipped with a thermo electric cooler to stabilize the output wavelength and comes in three wavelengths 405 nm 473 nm and 520 nm The DFB versions c...

Страница 6: ...ure 1 4 Using a flat blade screwdriver turn the Line Select Switch to the appropriate setting to match the AC input voltage you will be using 5 Remove the cover to the fuse holder You will find the fu...

Страница 7: ...tch to the appropriate position and install the proper fuse 2 Locate the unit on a dry level working surface 3 Make sure the POWER key switch on the front of the unit is in the OFF position key perpen...

Страница 8: ...er Cover should be installed on FC connector when fiber removed Display Select Switch Toggles the display between LD Power and temperature set point Keylock Power Switch Turn on power to the unit Key...

Страница 9: ...e end of your fiber optic cable may be less depending on the quality of the connection NOTE Each unit is calibrated internally to limit the maximum operating power of the laser diode to a safe operati...

Страница 10: ...is full counter clockwise the drive current is not zero but actually at the threshold current This should be kept in mind when using the modulation input 1 Connect a signal generator or 0 to 5 V power...

Страница 11: ...orlabs are configured with a shorting device installed in the Interlock connector If you are not going to use this feature then you can leave the shorting device installed and the unit will operate no...

Страница 12: ...the unit 7 2 Connector Cleaning Always clean the ferrule end of your fiber patch cable prior to inserting it into the output FC Adapter Your benchtop source comes with a fiber cleaning card FCC CLN4...

Страница 13: ...mbient Display Accuracy 10 Setpoint Resolution 0 01 mW Adjustment Range 0 mW to Full Power TEC Stability 0 005 C 1 C Setpoint Accuracy 0 25 C Setpoint Resolution 0 1 C Adjustment Range 20 1 C to 30 1...

Страница 14: ...Fiber Coupled Laser Source Chapter 9 Mechanical Drawing Page 12 19146 D02 Chapter 9 Mechanical Drawing Figure 6 Mechanical Drawing 12 49 317 3 mm 11 43 290 2 mm 2 59 65 7 mm 5 76 146 4 mm 3 06 77 8 mm...

Страница 15: ...C Certifocate of Compliance Rev L June 5 2017 Page 13 Chapter 10 FCC EMC Certifocate of Compliance Emissions comply with the Class A Limits of FCC Code of Federal Regulations 47 Part 15 Subpart B1 1 U...

Страница 16: ...orlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by...

Страница 17: ...ion USA Canada and South America Thorlabs Inc sales thorlabs com techsupport thorlabs com Europe Thorlabs GmbH europe thorlabs com France Thorlabs SAS sales fr thorlabs com Japan Thorlabs Japan Inc sa...

Страница 18: ...www thorlabs com...

Отзывы: