Fiber-Coupled Laser Source

 

Chapter 4: Setup

 

Page 4

 

18651-D02 

Chapter 4 

Setup 

4.1. 

Setting the AC Line Voltage and Installing Fuses 

Your  S1FC  Series  Laser  Source  has  been  shipped  from Thorlabs  configured  for  115  VAC  operation. If you  are 
planning to operate your unit using a 220/230 VAC input, or need to replace an open fuse, you must perform the 
following procedure. 

 

Remove the AC power cord if it is connected to the unit. 

 

Remove the cover of the unit by removing the two 4-40 phillips head screws located on the bottom rear of 
the unit and slide the cover off. Refer to Figure 1 on the following page. 

 

Locate the AC Line Select Switch and Fuse Holder. They are located near the AC Input Module towards 
the back of the unit. Refer to Figure 1. 

 

Using a flat blade screwdriver turn the Line Select Switch to the appropriate setting to match the AC input 
voltage you will be using. 

 

Remove  the  cover  to  the  fuse  holder.  You  will  find  the  fuse  installed  in  the  cover.  Remove  the  existing 
fuse and install the appropriate fuse for the line voltage you will be using: 

o

  For 115 VAC operation use 250 mA 

o

  For 220/230 VAC operation use 160 mA 

o

  In  all  cases  use  only  5  mm  x  20  mm  250  VAC  Type  T  Fuses  (IEC  60127-2/III,  low  breaking 

capacity, slow blow)  

 

Reinstall the cover and replace the two 4-40 phillips head screws. 

 

Remove the small cap head screw indicating the previous voltage configuration located on the rear panel, 
above  the  AC  input  housing.  Place  the  screw  in  the  threaded  hole  indicating  the  new  AC  input 
configuration, see Figure 2. 

4.2. 

Initial Set-up 

 

Determine the AC line voltage the unit will be connected to (either 115 VAC or 230 VAC) and set the AC 
Line Voltage Selector Switch to the appropriate position and install the proper fuse.  

 

Locate the unit on a dry, level working surface. 

 

Make  sure  the  POWER  key  switch  on  the  front  of  the  unit  is  in  the  OFF  position  (key  perpendicular  to 
working surface). 

 

Plug  the  female  end  of  the  AC  line  cord  provided  into  the  AC  Input  Receptacle  on  the  rear  of  the  unit. 
Plug the male end into a properly grounded AC socket. 

 

Connect a Fiber Optic cable to the LASER APERTURE on the front panel of the unit. 

Содержание S1FC1310PM

Страница 1: ...S1FC635PM S1FC780PM S1FC1310PM S1FC1550PM Fiber Coupled Laser Source User Guide ...

Страница 2: ...ront and Back Panel Overview 6 5 2 Turning on the Source 7 5 3 Adjusting the Laser Output Power 7 5 4 Turning the Laser Off 7 5 5 Modulating the Laser Output 7 Chapter 6 Making the Safety Interlock Connection 8 Chapter 7 General Maintenance 9 7 1 Cleaning 9 7 2 Connector Cleaning 9 Chapter 8 Specifications 10 Chapter 9 Mechanical Drawing 11 Chapter 10 Regulatory 12 Chapter 11 Certificate of Compli...

Страница 3: ...mbol Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off Supply In Position of a Bi Stable Push Control Out Position of a Bi Stable Push Control Caution Risk of Electric Shock Caution Hot Surface Caution Risk of Danger Warning Laser Radiation Caution Spinning ...

Страница 4: ...o not obstruct the air ventilation slots in the housing Make sure that the line voltage rating marked on the rear panel agrees with your local supply and that the appropriate fuses are installed Changing of the mains fuse can be done by the user see Setting the AC Line Voltage and Installing Fuses With the exception of the mains fuses there are no user serviceable parts in this product This device...

Страница 5: ...version wavelength choices from 405 to 1550 nm 635 nm to 1550 nm for PM versions with standard single mode fiber or polarization maintaining fiber output The S1FCxPM Series comes with a pigtailed Fabry Perot Laser diode with polarization maintaining fiber behind an FC PC bulkhead connector with the key aligned to the slow axis The DFB version comes equipped with a thermo electric cooler to stabili...

Страница 6: ...he existing fuse and install the appropriate fuse for the line voltage you will be using o For 115 VAC operation use 250 mA o For 220 230 VAC operation use 160 mA o In all cases use only 5 mm x 20 mm 250 VAC Type T Fuses IEC 60127 2 III low breaking capacity slow blow Reinstall the cover and replace the two 4 40 phillips head screws Remove the small cap head screw indicating the previous voltage c...

Страница 7: ...upled Laser Source Chapter 4 Setup Rev E February 21 2017 Page 5 Figure 1 Locations of Line Select Switch and Fuse Holder Line Select Switch Set to either 110 or 220 V Fuse Holder Fuse is installed in the cover ...

Страница 8: ...onnector Cover Screws 2 Remove to access line voltage switch and fuse Laser Output Aperture Cover Cover should be installed on FC connector when fiber removed Laser Enable Switch Press to activate laser Laser Emission Indicator Lights up 5 seconds before laser turns on Keylock Power Switch Turn on power to the unit Key only removes when off Power Adjust Knob Remote Interlock Input Modulation Input...

Страница 9: ...l turn OFF the entire unit Anytime the unit is turned OFF and then turned back ON the Laser will be disabled until the ENABLE switch is pressed 5 5 Modulating the Laser Output The MOD IN input can be used to modulate the laser output or set the laser output remotely using a 5 V power source The 5 V maximum input corresponds to the maximum calibrated power of the laser source Each unit is calibrate...

Страница 10: ...bs are configured with a shorting device installed in the Interlock connector If you are not going to use this feature then you can leave the shorting device installed and the unit will operate normally as described in the procedures above If you wish to make use of the Interlock feature you will need to acquire the appropriate connector mate and wire it your remote interlock switch Next remove th...

Страница 11: ...or near the unit 7 2 Connector Cleaning Always clean the ferrule end of your fiber patch cable prior to inserting it into the output FC Adapter Your benchtop source comes with a fiber cleaning card FCC CLN2 1 This should be used before inserting the fiber connector into the mating barrel Figure 4 Fiber Cleaning Card FCC CLN2 1 To use the card peel back and tear away one small blue strip Holding th...

Страница 12: ...curacy 10 Setpoint Resolution 0 01 mW Adjustment Range 0 mW to Full Power Fiber Alignment Slow Axis Aligned to the Key TEC Stability n a Setpoint Accuracy n a Setpoint Resolution n a Adjustment Range n a Environmental Operating Temperature 15 to 35 C Storage Temperature 0 to 50 C AC Input 115 VAC 230 VAC Switch Selectable 50 60 Hz Modulation Input 0 5 V 0 Full Power DC or Sine Wave Input Only Modu...

Страница 13: ...Fiber Coupled Laser Source Chapter 9 Mechanical Drawing Rev E February 21 2017 Page 11 Chapter 9 Mechanical Drawing ...

Страница 14: ...horlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by the user PCB s housings etc If you wish to return a Thorlabs unit for waste recovery please contact Thorlabs or your nearest dealer for further information Waste Treatment is Your Own Responsibility ...

Страница 15: ...te of Compliance Rev E February 21 2017 Page 13 Chapter 11 Certificate of Compliance FCC EMC Compliance Emissions comply with the Class A Limits of FCC Code of Federal Regulations 47 Part 15 Subpart B1 1 Using high quality shielded interface cables ...

Страница 16: ... USA Canada and South America Thorlabs Inc sales thorlabs com techsupport thorlabs com Europe Thorlabs GmbH europe thorlabs com France Thorlabs SAS sales fr thorlabs com Japan Thorlabs Japan Inc sales thorlabs jp UK and Ireland Thorlabs Ltd sales uk thorlabs com techsupport uk thorlabs com Scandinavia Thorlabs Sweden AB scandinavia thorlabs com Brazil Thorlabs Vendas de Fotônicos Ltda brasil thorl...

Страница 17: ......

Страница 18: ...www thorlabs com ...

Отзывы: