PMT2100 Series Photomultiplier Tubes
Chapter 6: Software
Page 26
TTN094969-D02
3. Click the
button next to each PMT to expand the respective PMT
panel.
Figure 39 PMT2100 Control Window
Note:
If the serial number is unavailable, select the respective PMT on the
PMT2100 Control
window, and click the
Update Settings SN
button to update the
PMT’s serial number in the
PMTSettings
file.
6.5.
Setting the PMT2100 Offset
You must set the offset for each of the PMT2100 PMT devices.
Note:
For a Galvo-Resonant system, once the initial offset adjustment is set, you
do not have to adjust the offset for every use. For a Galvo-Galvo system, you may
need to adjust the offset depending on the Dwell Time setting.
1. Switch off the room lights.
Note:
An unpowered PMT exposed to bright light can have a high dark
count that may take several minutes to hours to dissipate.
2. Open the
PMT2100 Control
window.
3. Use the table below to configure all the PMT2100s.
Scanner
PMT Input
Range
Bandwidth
Gain
Galvo-Resonant
1 V
80 MHz
0
Galvo-Galvo
1 V
Large Dwell Time
(Slow Scan): 250 kHz
Small Dwell Time (Fast
Scan): 2.5 MHz
0
4. Capture an image on your image scanning equipment.
5. Adjust the gain on
PMT2100 Control
window until the average of all the
pixels in the image is 50 - 100.