3 Operating Instruction

7

P

opt,1

 and P

opt,2

 - optical input power

G - transimpedance gain of the RF output 

The responsivity 

Â

(

l

)

 for a given wavelength can be read from the individual curves in section

Technical  Data  to  estimate  the 

RF  OUTPUT

  voltage.  Please  note  that  the  given  responsivity

curves represent typical values - individual responsivity may deviate.    

The maximum 

RF OUTPUT power 

is typ 18 dBm at 1 dB compression. To explain that:

An  ideal  amplifier  would  be  a  linear  device  -  it's  gain  is  constant  at  any  input  power  level.  In
reality, the output power of the amplifier is limited (e.g. by supply voltage). That means, with in-
creased  input  power  the  amplifier  becomes  saturated.  Before  the  saturation  is  reached,  a  re-
duction in gain takes place - the result is a compression effect. The 1 dB compression point is
defined as the output power where the gain is 1 dB less than the small signal gain.

In  order  to  avoid  saturation,  the  output  signal  level  should  not  exceed  this  1-dB-compression
point.

MONITOR Outputs

The signal monitor outputs (

 and 

MONITOR-

) allow observation of the input power

level and can be used as individual power indicators. These outputs can also be used to meas-
ure  RF  (up  to  3  MHz)  modulated  signals  on  each  detector  separately.  The  maximum  output
voltage  swing  of  the 

MONITOR

  output  is  +10 V  for  high  impedance  loads  (+1.5 V  into  50 

W

).

Saturation of 

MONITOR

 outputs will occur at optical input power level greater than 1 mW, de-

pending on the detector's wavelength response.

The amplifier offset voltage is factory set to zero at 23°C ambient temperature. A small drift dur-
ing a short warm-up period (~5min) may occur. For exact DC light level measurements a con-
stant temperature environment is recommended.

3.4   Mounting

The PDB48xC-AC is housed in a rugged shielded aluminum enclosure. 

For post mounting an adapter can be attached to the bottom or side surface using four M2x8
screws (see below). This adapter supports #8-32 as well as M4 post mounts. The M4 tread is
marked.

    

Содержание PDB48*C-AC series

Страница 1: ...Balanced Amplified Photodetectors PDB48xC AC Operation Manual 2019...

Страница 2: ...Version Date Item No 1 6 14 Jan 2019 M 0009 510 666 Copyright 2019 Thorlabs...

Страница 3: ...Inputs 6 3 3 Electrical Outputs 7 3 4 Mounting 8 3 5 CMRR and Frequency Response 9 3 6 Recommendations 4 Maintenance and Service 10 5 Appendix 11 12 5 1 Technical Data 13 5 1 1 Individual Diagrams PDB...

Страница 4: ...nternational partners are looking forward to hearing from you Thorlabs GmbH Warning Sections marked by this symbol explain dangers that might result in personal injury or death Always read the associa...

Страница 5: ...s contain detailed information about principle of operation operating suggestions and technical specifications 1 1 Safety Attention The safety of any system incorporating the equipment is the responsi...

Страница 6: ...ange on the power supply matches with your local mains voltage If you want use your own power supply Thorlabs offers an ap propriate power connector cable Carefully unpack the unit and accessories If...

Страница 7: ...or other low frequency measurement device After finishing measurements turn power off Attention The damage threshold of the photo diodes is 10 mW PDB480C AC resp 5 mW PDB481C AC Exceeding this value w...

Страница 8: ...level should not exceed the RF output power 1 dB compression point Attention The damage threshold of the photo diodes is 10 mW PDB480C AC resp 5 mW PDB481C AC Exceeding this value will permanently de...

Страница 9: ...nal gain In order to avoid saturation the output signal level should not exceed this 1 dB compression point MONITOR Outputs The signal monitor outputs MONITOR and MONITOR allow observation of the inpu...

Страница 10: ...DuT The output of the DuT that should be measured is connected to the network analyzers Port B Frequency response measurements The frequency response of each signal path can be measured by connecting...

Страница 11: ...the two signals which will decrease the noise reduction capability of the balanced detector The figure below shows the maximum allowed path length difference in fiber to obtain a de sired CMRR Avoid e...

Страница 12: ...ectors The photodiodes of the PDB48xC AC are pigtailed with single mode fiber and connectorized by FC APC con nectors Clean and dust free surfaces of the ferrule tips are essential to minimize couplin...

Страница 13: ...red across a 50 W load at large bandwidth e g if connect the RF output to a 50 W terminated scope input Max input power is the damage threshold of the photo diode Typical noise spectra diagrams These...

Страница 14: ...er at 1 dB compression 50 W load min 16 5 dBm typ 18 dBm RF OUTPUT Coupling AC coupling only RF Output Impedance 50 W Minimum NEP 30 kHz to 100 MHz Overall Output Voltage Noise 9 mVRMS 6 5 mVRMS MONIT...

Страница 15: ...2019 Thorlabs 5 Appendix 13 5 1 1 Individual Diagrams PDB480C AC PDB480C AC Typical Detector Responsivity PDB480C AC Typical RF OUTPUT Frequency Response...

Страница 16: ...2019 Thorlabs 14 PDB48xC AC PDB480C AC Typical MONITOR Output Frequency Response PDB480C AC Typical RF OUTPUT Spectral Noise...

Страница 17: ...2019 Thorlabs 5 Appendix 15 5 1 2 Individual Diagrams PDB481C AC PDB481C AC Typical Detector Responsivity PDB481C AC Typical RF OUTPUT Frequency Response...

Страница 18: ...2019 Thorlabs 16 PDB48xC AC PDB481C AC Typical MONITOR OUTPUT Frequency Response PDB481C AC Typical RF OUTPUT Spectral Noise...

Страница 19: ...2019 Thorlabs 5 Appendix 17 5 2 Dimensions PDB48xC AC Mechanical Drawing...

Страница 20: ...2019 Thorlabs 18 PDB48xC AC 5 3 Certifications and Compliances...

Страница 21: ...l or the technical data of the de scribed unit at any time 5 5 Copyright and Exclusion of Reliability Thorlabs has taken every possible care in preparing this document We however assume no li ability...

Страница 22: ...Brazil Tel 55 16 3413 7062 Fax 55 16 3413 7064 www thorlabs com Email brasil thorlabs com Japan Thorlabs Japan Inc 3 6 3 Kitamachi Nerima ku Tokyo 179 0081 Japan Tel 81 3 6915 7701 Fax 81 3 6915 7716...

Страница 23: ...www thorlabs com...

Отзывы: