![THORLABS PCS-6000 Series Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/thorlabs/pcs-6000-series/pcs-6000-series_user-manual_1110773019.webp)
Rev B
,
January 10, 2020
Page 15
Chapter 5 Installation
5.1.
Unpacking the Unit
Carefully unpack the unit and inspect for shipping damage. Severe damage to
the packaging may indicate component damage. Notify your shipping/receiving
department immediately.
5.2.
Removing the Shipping Clips
To prevent damage to the stages and piezo during shipment, the stage top and
bottom are displaced from their normal positions with shipping clips. Due to ten-
sions, these shipping clips must be removed carefully in order to prevent dam-
age. Follow the steps below to remove the clips:
1. Use the 5/64" hex driver (red handle) to remove the screw going through
the thicker portion of the clip.
Figure 5–1 Removing the Bottom Screw
2. Gently push the moveable portion of the stage to create a ~1 mm gap
between the thicker portion of the clip and the stationary stage member.
Figure 5–2 Creating the Gap
3. Remove the top screw and gently allow the stage to slide into its rest
position.
4. Repeat steps 1-3 for the remaining shipping clips.
5.3.
Mounting the Manipulator
1. Locate the rotation plate adapter and determine the approximate
location where it is to be mounted.
PCS-6000 Series Motorized Patch-Clamp Micromanipulator
Chapter 5: Installation