![THORLABS OCT-LK Series Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/thorlabs/oct-lk-series/oct-lk-series_user-manual_1110880014.webp)
OCTP Series Scanner
Chapter 3: Components and Compatibility
Rev F, November 03, 2018
Page 9
3.1.6. Lens Kits
For the case that an OCT scan lens kit is installed, the mechanical dimensions are summarized in the table
below.
Figure 8 Objective Holder Shown with Scan Lens Kit Installed
OCT Scan Lens Kit
OCT-LK2(-BB)
OCT-LK3(-BB)
OCT-LK4(-BB)
Field of View
6 mm x 6 mm
10 mm x 10 mm
16 mm x 16 mm
Barrel Height (B)
29.5 mm
28.0 mm
41.0 mm
Working Distance (WD)
3.4 mm
24.9 mm
41.6 mm
Table 1 Dimensions for OCT-LKx Scan Lens Kits
Changing the objective of the OCTP requires changing the probe configuration data in the software package
ThorImage
®
OCT. The length of the reference arm must be aligned to match the optical path in the sample arm
as well.
This is performed automatically using the built-in calibration procedure of the ThorImage
®
OCT software package
version 4.1.4 and higher. A detailed description of this procedure is given in the ThorImage
®
OCT operating
manual.
3.1.7. OCT-STAND
The OCT-STAND is a dedicated stand for OCTP scanners. The OCTP scanners are attached to the focus block
of the OCT-STAND using a spring-loaded mount accepting the dovetail mount of the OCTP. The focus block
can be rotated 360° around the Ø1.5” rod, and features 30 mm of travel with fine and coarse adjustment knobs.
Figure 9 OCTP in Thorlabs OCT-STAND
Содержание OCT-LK Series