Swept Source OCT System
Chapter 1: Introduction
Rev H, June 3, 2011
Page 7
Figure 1
Fiber Inspection Scope (FS200)
Please follow the steps below to inspect the fiber end:
Ensure that any source of Laser Radiation is TURNED OFF. This OCT system provides both IR and visible laser
radiation, so both sources must be disabled. The lasers are off
when the “LASER ON” indicator on the front panel
of the Swept Source Engine is not illuminated
, and when the “AIM ON” indicator on the Imaging Module is not
illuminated.
1. Insert the terminated fiber into the fiber input.
2. Look into the eyepiece and press the on/off switch.
3. Adjust the focus control to find the clearest image.
4. Rotate the fiber end to make sure that no dirt is rotating in the view. Dirt that rotates with the fiber is
attached to the tip of the ferrule.
5.
If the fiber tip is dirty, draw figure 8’s with the ferrule on the cleaning reel. Re-inspect, and repeat if
necessary.
1.4.2. Service
Only tr
ained and approved Thorlabs’ personnel should service the system. Please contact Thorlabs’ Technical
Support at
and a member of our team will be happy to assist you.
1.4.3. Accessories and Customization
Though the system is easily adapted for custom interfaces, to achieve the listed specifications, this system should
only be used with accessories provided by Thorlabs. Any modification or servicing by unqualified personnel
renders the warranty null and void, leaving Thorlabs free of liability. Please contact Thorlabs for questions on
customization.
Focus adjustment
Fiber input
Eyepiece
On/off switch