Chapter 3 Installation

3.4 Dimensions

3.4.1 MLS203-1 Dimensions

Fig. 3.5   MLS203-1 Dimensions

140.0mm

5.51in

115.0mm

4.53in

79.0mm

3.11in

54.0mm

2.13in

172.0mm

6.77in

72.0mm

2.83in

70.0mm

2.76in

112mm

4.4

APPROX

260.0mm

10.24in

120.0mm

4.72in

191.0mm

7.52in

170.5mm

6.71in

160.3mm

6.31in

214.0mm

8.43in

130.5mm

5.14in

125.0mm

4.92in

120.3mm

4.74in

58.5mm

2.30in

20.5mm

0.81in

M3X0.5 TAPPED HOLE

4 PLACES

M3X0.5 TAPPED HOLE FOR

MOUNTING MLS203P1-5 

ACCESSORY PLATES 4 PLACES
OBJECTIVE CENTRE AND

CENTRE OF STAGE TRAVEL
MOUNTING HOLE FOR M5 

CAP SCREW 4 PLACES

M3 x 0.5 TAPPED HOLE 

6 PLACES

REMOVABLE GUARD

REMOVABLE GUARD

31.1mm

1.22in

10.0mm

0.39in

230.0mm

9.06in

X AXIS MOTOR AND

ENCODER CABLES

Y AXIS MOTOR AND

ENCODER CABLES

25.0mm

0.98in

ACCESSORY PLATE MOUNTING SURFACE

All dimensions in mm [in.]

Содержание MLS203-1

Страница 1: ...HA0264T MLS203 1 Fast XY Scanning Stage User Guide Original Instructions...

Страница 2: ...onnections 6 3 4 Dimensions 7 3 4 1 MLS203 1 Dimensions 7 Chaper 4 Operation 8 4 1 General 8 4 2 Using the Kinesis software 9 4 3 Using the APT software 12 4 4 Position Error Messages 14 4 5 Transport...

Страница 3: ...on uses high rigidity precision linear bearings commercial grade limit switches and powerful brushless linear DC servo motors In addition the number of moving parts has been minimized which optimizes...

Страница 4: ...sk of Electrical Shock Given when there is a risk of injury from electrical shock Warning Given when there is a risk of injury to users Caution Given when there is a risk of damage to the product Note...

Страница 5: ...shims which should be fitted as necessary for more details see the instructions for fitting the attachment brackets When mounting the stage close to other equipment ensure that the travel of the movi...

Страница 6: ...er guards to avoid trapping of body parts or cables in the mechanism These should be fitted before the stage is mounted on the microscope as follows 1 Offer up the finger guards to the stage as shown...

Страница 7: ...e the manual XY specimen stage 3 Place the stage onto the mounting platform and loosely fit the three M4 x 12 bolts provided 4 Check that the mounting surface is flat and that the stage does not rock...

Страница 8: ...e correct channel Fig 3 3 Electrical connections Pin out information for the motor drive and encoder feedback connectors on the motor flying leads is detailed below Fig 3 4 Motor Drive and Feedback Fl...

Страница 9: ...n 214 0mm 8 43in 130 5mm 5 14in 125 0mm 4 92in 120 3mm 4 74in 58 5mm 2 30in 20 5mm 0 81in M3X0 5TAPPED HOLE 4 PLACES M3X0 5TAPPED HOLE FOR MOUNTING MLS203P1 5 ACCESSORY PLATES 4 PLACES OBJECTIVE CENTR...

Страница 10: ...5 If no joystick is being used click the Home button on the GUI panel Caution The MLS203 stages is designed to be driven by the Thorlabs BBD20x or RBD201 Brushless DC Motor Controllers Warning The mot...

Страница 11: ...set up parameters When the stage is connected these settings are loaded into the controller on start up and are tuned for loads up to the 1 0 kg 2 2 lb maximum at speeds up to 250 mm s However dependi...

Страница 12: ...tune the control loop for your application see Table 4 1 and Table 4 2 for more information 4 After the parameter changes have been performed click the Persist Settings to Hardware box then click OK...

Страница 13: ...s are mounted in an XY configuration Table 4 2 Load vs Maximum acceleration recommendations Load g Approximate Max Acceleration mm s 0 5000 125 2400 250 1550 500 925 750 650 900 500 Note Position PID...

Страница 14: ...et up parameters When the stage is connected these settings are loaded into the controller on start up and are tuned for loads up to the 1 0 kg 2 2 lb maximum at speeds up to 250 mm s However dependin...

Страница 15: ...ave been performed click the Persist Settings to Hardware box then click OK This will ensure that the same parameter settings will be loaded next time the unit is powered up even in the absence of a P...

Страница 16: ...n the stage is controlled by the BBD20x controller the maximum velocity is limited to safe values However if the output is disabled with the controller connected and monitoring the position and the st...

Страница 17: ...e information 4 7 Transportation Caution When packing the unit for shipping use the original packing If this is not available use a strong box and surround the unit with at least 100 mm of shock absor...

Страница 18: ...near Bearing Motor Type Brushless DC Linear Motor Dimensions Mid Travel 10 24 x 9 06 x 1 22 260 0 mm x 230 0 mm x 31 1 mm Weight with cables 7 0 lbs 3 2 kg Note The default tuning parameters can be ch...

Страница 19: ...microscope sample platforms has been designed to provide flexible options of sample types traditional slides petri dishes or well plates as well as uniquely shaped sample holders for biofilms microflu...

Страница 20: ...st harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with...

Страница 21: ...Rev F Apr 2019 Page 19 Chapter 7 Regulatory 7 2 CE Certificates...

Страница 22: ...Thorlabs for more information Waste treatment is your own responsibility End of life units must be returned to Thorlabs or handed to a company specializing in waste recovery Do not dispose of the unit...

Страница 23: ...www thorlabs com...

Отзывы: