THORLABS LDM21 Скачать руководство пользователя страница 7

LDM21

 

Chapter 3: Setup 

TTN013394-D02

 

 

Page 4

 

1.  The four sockets comprising the laser diode connector are through-hole 

type  sockets  with  a  blind  clearance  of  0.48”  measured  from  the  front 
face of the copper cold plate. It is not necessary to trim the laser diode 
leads prior to mounting into this connector unless they are longer than 
0.48”. 

2.  The  laser  connector  is  located  close  enough  to  the  front  face  of  the 

copper  cold  plate  to  allow  easy  installation  of  short-leaded  lasers.  The 
clearance  area  around  the  LD  and  PD  sockets  is  sufficient  to  prevent 
the pins from contacting the cold plate. 

3.  Most laser diodes are three pins  with the case tied to one of the laser 

pins  and  also  to  one  of  the  photodiode  pins.  The  other  laser  and 
photodiode  pins  will  be  isolated  from  the  case.  The  LDM21  was 
designed  to  operate  the  laser  case  at  ground  potential,  therefore  this 
common  pin  will  be  inserted  into  either  the  12  o’clock  or  the  6  o’clock 
position of the laser connector. Locate the isolated laser pin and insert it 
in the 3 o’clock position. The isolated photodiode should now be in the 9 
o’clock position. Refer to Fig. 3. (Orient the mount with the PD polarity 
switch above the LD socket). 

4.  The  LDM21  Mount  is  also  compatible  with  style  E  and  style  H  laser 

diodes,  which  do  not  have  a  photodiode;  they  have  only  a  laser  diode 
and a ground pin. They can therefore be set up the same as style A (LD 
Anode Ground) or style B (LD Cathode Ground) respectively. 

5.  Replace  the  laser  mounting  flange  and  the  cover.  Install  both  screws 

through  the  mounting  flange  and  loosely  into  the  mount.  Carefully 
tighten each screw a little bit at a time until the flange is just snug. Do 
not  over-tighten  either  screw-  the  flange  should  sit  slightly  above  the 
cold  plate.  Reinstall  the  cover  using  the  four  2-56  cap  head  screws 
provided. 

 

Figure 3  LD and PD Orientation 

 

G

G

LD

PD

Содержание LDM21

Страница 1: ...LDM21 5 6 mm 9 mm Laser Diode Mount User Guide...

Страница 2: ......

Страница 3: ...5 3 5 TEC Controller Connections 6 3 6 Thermistor Data 7 3 7 Mounting Other Accessories 9 Chapter 4 Operation 10 4 1 Safe Operating Area 10 4 2 Maintaining the LDM21 11 Chapter 5 Troubleshooting 12 C...

Страница 4: ...ption Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off Supply In Posi...

Страница 5: ...mount are 4 40 tapped holes on 30 mm centers for mounting any number of Thorlabs cage assembly products There is also a 8 32 tapped hole on the bottom surface to allow the unit to be post mounted The...

Страница 6: ...2 56 screws from the laser mounting flange and remove the flange 3 Determine the laser pin configuration from the laser diode manufacturer s data sheets and set the LD Laser Diode and PD Photodiode s...

Страница 7: ...case at ground potential therefore this common pin will be inserted into either the 12 o clock or the 6 o clock position of the laser connector Locate the isolated laser pin and insert it in the 3 o...

Страница 8: ...The LDM21 is compatible with all Thorlabs LDC LD controllers and ITC series combination controllers LD and TEC Appropriate cables with DB9 connectors are included with Thorlabs controllers and ensure...

Страница 9: ...socket when the PD Polarity Switch is set to AG Anode Grounded 5 Interlock Return This pin is the return side of the Interlock circuitry Pin 1 and 5 are shorted internally in the LDM21 6 N A 7 Laser...

Страница 10: ...turn pin 4 TEC This pin is connected to the positive terminal of the TEC element 5 TEC This pin is connected to the negativeterminal of the TEC element 6 N A 7 AD592 The negative terminal of the AD592...

Страница 11: ...the resistance value at a reference temperature which is room temperature in this case T is the temperature in K Constants Value Unit A 14 6337 B 4791 842 K C 115334 K2 D 3 730535E 06 K3 A1 3 354016E...

Страница 12: ...s mounts This is most often used for mounting aspheric collimating optics available separately from Thorlabs Also included are four 4 40 tapped holes mounted on 30 mm centers for attaching our cage as...

Страница 13: ...s intended to be operated at a fixed temperature set point of 25 C but can be operated as low as 20 C as long as the input power to the laser diode input current x LD voltage does not exceed the value...

Страница 14: ...leaned by wiping with a soft damp cloth If you suspect a problem with your LDM21 please call Thorlabs Technical Support and an engineer will be happy to assist you 0 100 200 300 400 500 600 700 20 22...

Страница 15: ...the rear panel of your LDC series laser diode controller The laser diode is installed into the wrong pins on the laser diode socket Refer to Fig 2 see page 5 for the correct orientation of the laser...

Страница 16: ...8 32 and M4 Threaded Mounting Holes Laser Specifications Lasers Supported 5 6 mm and 9 mm Max Laser Current 500 mA1 Max Laser Input Power 600 mW1 Laser Pin All LD Packages Switch Selectable Configurat...

Страница 17: ...ical Drawing Figure 5 Mechanical Drawing of the LDM21 1 75 1 75 0 88 0 88 30MM CAGEASSEMBLY MOUNTING HOLES 4 40 4 PLACES 0 88 0 12 TEC CONNECTION MALE DB9 1 66 0 88 0 12 LASER DIODE CONNECTION FEMALE...

Страница 18: ...roducts such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by the user...

Страница 19: ...nd South America Thorlabs Inc sales thorlabs com le techsupport thorlabs com Europe Thorlabs GmbH europe thorlabs com France Thorlabs SAS SA sales fr thorlabs com Japan Thorlabs Japan Inc sales thorla...

Страница 20: ...www thorlabs com...

Отзывы: