2 Getting Started Quickly
INT-POL-1300
/ page 4
Switch the power supply to your local main voltage (100 VAC, 120 VAC or
230 VAC), see Figure 1.
Figure 1:
Switchable power supply for 100
V, 120 V, or 230
V
Plug the power connector cable into the
DC INPUT
.
Plug the power supply into a 50-60 Hz, 100 VAC outlet (120 VAC, or 230 VAC
resp.) and turn it on.
Connect
RF
OUTPUTs
and, if required,
MONITOR
outputs with coaxial cables
to your data acquisition device.
2.3 First Operation
Please note, that all fiber connectors must have a FC/APC connector!
Connect the INT-POL-1300 to your OCT setup, for further details please refer
to section 3.4 Basic Measurement Setup.
For balanced operation (optimum noise suppression), the power at each de-
tector pair has to be the same. This can be achieved by inserting polarization
controllers previous both optical inputs and change the SOP so that both di-
odes are balanced. The monitor outputs helps to adjust the power to the same
level.
The maximum
RF-OUTPUT
voltage swing is
3.6 V for high impedance loads
(
1.8 V into 50
loads). The
RF OUTPUT
signal should not exceed this max-
imum output voltage to avoid saturation. If necessary, reduce optical input
power at input
ports.
The maximum
MONITOR
voltage swing is +10 V for high impedance loads
(+1.5 V into 50
loads). The
MONITOR
signal should not exceed this maxi-
mum output voltage to avoid saturation. If necessary, reduce optical input
power at input
ports.