HNLS008R(L) Red HeNe Laser System
Chapter 9: Compliance
Rev C, March 23, 2015
Page 18
Chapter 9
Compliance
FDA-CDRH Compliance
Under the US FDA CDRH, the lasers conform to the Code of Federal Regulations (CFR), Title 21, Subchapter J,
which pertains to laser safety and labeling. See www.fda.gov/cdrh/radhlth/cfr/21cfr1000-1050.pdf for more
information.
If the end user integrates this equipment into another product, it is the customer’s responsibility to ensure CDRH
certification of their instrument.
CE Compliance
The HNL Series lasers comply with the following standards:
IEC/EN 60825-1 Concerning Laser Safety
IEC/EN 61010-1 Concerning Electrical Safety
EN61326 Concerning EMC
The HNL Series power supplies comply with the following standards:
IEC/EN 61010-1 Concerning Electrical Safety
EN61326 Concerning EMC
Conditions of Acceptability
When installed in the end product, consideration shall be given to the following:
A suitable fire enclosure shall be provided in the end product.
Safety interlock switch, key switch, laser housing and laser beam attenuator in accordance with laser
safety standards shall be provided in the end product.
A visual or audio indicator, in accordance with laser safety standards, shall be provided in the end
product.
The unit's thermal circuitry shall be evaluated in the end product.
Compliance with FDA or other laser safety standards shall be determined in the end-use application.
The end user must provide his or her own safety monitoring mechanism to shut down a power supply if it
fails to start the laser after several seconds.
The end user must use laser power supplies that provide a protective earth ground wire connected to the
mains ground. This wire must be able to support 25 A DC or AC RMS for one minute.
The HNL Series require that the power supply high-voltage negative output be connected to earth ground.
Содержание HNLS008L
Страница 1: ...HNLS008R HNLS008L Self Contained Red HeNe Laser System User Guide...
Страница 20: ...www thorlabs com...