Page 10
18728-D01
Single- and Dual-Axis Scanning Galvo Systems for Small Beam Diameters
3.2 Mechanical Installation
3.2.1 Introduction
.
Fig. 3.7 Serial Number Label Location
It is essential that the user mounts heatsinks to the driver board and motor mounts
which are suitable for their intended application. If this is not done the devices will
overheat and permanent damage may occur. The choice of heatsink will primarily be
determined by the power which the devices dissipate, a value which is dependant on
the average speed at which the user moves the scanners. The larger the power the
heatsink must dissipate the larger the heatsink will need to be.
3.2.2 Fitting The Heatsinks
Caution
The galvo motor assembly and associated driver board are tuned at the factory
before they are shipped and further adjustment is not normally necessary. If the
accuracy of the system is in doubt, e.g due to accidental adjustment of trim pots,
contact Thorlabs for information on the tuning procedure.
During Installation, ensure that the motors are connected to the driver card to
which they were tuned. Both the motor and the driver card should carry the
same serial number.
The location of the serial number labels is shown below.
Содержание GVS001
Страница 33: ...Rev 22 Feb 2019 Page 31 Chapter 5 Troubleshooting D 2 CE Certificates ...
Страница 34: ...Page 32 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Страница 35: ...Rev 22 Feb 2019 Page 33 Chapter 5 Troubleshooting ...
Страница 36: ...Page 34 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Страница 37: ...Rev 22 Feb 2019 Page 35 Chapter 5 Troubleshooting ...
Страница 38: ...Page 36 18728 D01 Single and Dual Axis Scanning Galvo Systems for Small Beam Diameters ...
Страница 39: ...Rev 22 Feb 2019 Page 37 Chapter 5 Troubleshooting ...
Страница 41: ...www thorlabs com ...