6 Maintenance and Service
Protect the DMH40 from adverse weather conditions. The DMH40 is not water resistant.
Attention
To avoid damage to the instrument, do not expose it to spray, liquids or solvents!
The unit does not need regular maintenance by the user. It does not contain any modules
and/or components that could be repaired by the user himself.
Do not open the housing!
Since the optical mirror of the DMH40 is very thin and mounted on a flexible surface, it is ex-
tremely vulnerable!
Attention
The DMH40 must not be cleaned from dust using ethanol, cleaning tissue, cotton tipped applic-
ators or any other mechanical tools! Using these chemical and / or mechanical tools voids war-
ranty!
Remove dust using only oil- and water-free compressed gas, such as Thorlabs Duster
(Tetrafluoroetane). Keep the gas nozzle at least at a distance of 4 inches (10 cm)
from the mirror surface, otherwise liquid gas drops may hit the surface and leave visible traces
on the mirror surface; moreover the air pressure might damage the mirror.
In case of malfunction of the DMH40, please contact
.
6.1 Version Information
Select the
icon from the
HELP
menu or in the toolbar for detailed information about the soft-
ware and the recognized DMH40:
The keyboard button
F1
brings up the online help - either the entire help document or context
sensitive.
Note
58
53
Содержание DMH40-F01
Страница 1: ...Piezoelectric Deformable Mirror 2021 DMH40 M F01 DMH40 M P01 Operation Manual...
Страница 2: ...Version Date 1 1 15 Apr 2021 Copyright 2021 Thorlabs...
Страница 56: ...2021 Thorlabs 51 DMH40 7 4 Dimensions DMH40 xxx Imperial...
Страница 59: ...2021 Thorlabs 7 Appendix 54 7 8 Certifications and Compliances...
Страница 66: ...www thorlabs com...