4 Computer Interface
DC3100
31
4.3.1.1.24 Revision Query
Command:
DC3100_revisionQuery
Parameter:
None
Response:
Instrument driver revision
Firmware revision
Description:
This function returns the instruments driver revision and the
devices firmware revision.
4.4
Status Reporting
The DC3100 stores the status in a register. It can be accessed via the
'DC3100_getStatusRegister. The following table lists all status numbers and the
according descriptive messages. Each bit represents an error.
Status
Bit
Name
Description
Bit 0
No LED Changed
The bit 'No LED' has changed.
Bit 1
No LED
The LED head is not connected to the chassis.
Bit 2
VCC Fail Changed
The bit 'VCC Fail' has changed.
Bit 3
VCC Fail
The power supply is out of range.
Bit 4
OTP Change
The bit 'OTP' (Over Temperature) has
changed.
Bit 5
OTP
Over temperature (OTP) in the chassis was
detected. The LED head was switched off.
Bit 6
LED Open Changed
The bit 'LED Open' has changed.
Bit 7
LED OPEN
The LED head is connected but no LED was
installed.
Bit 8
Limit Changed
The bit 'Limit' has changed.
Bit 9
Limit
The adjusted current is greater than the current
limit and was coerced to the limit.
Bit 10
OTP Head Changed
The bit 'OTP Head' (Over Temperature Head)
has changed.
Bit 11
OTP Head
Over temperature in the LED head was
detected. The LED head was switched off.
Bit 12
Interface Refresh
The user has changed settings.