Compact Scientific Digital Cameras
Chapter 3: Setup & Installation
Page 16
ITN004347-D02
3.6.
Optical Front End Procedures
3.6.1. Window Removal and Replacement Procedure
TSI Compact Scientific cameras are provided with a protective glass window installed (See Chapter 7).
The optical front end is designed to accommodate standard Ø1" windows. Users may, at their discretion, operate
the camera without any glass installed, or install custom filters in front of the sensor. The maximum filter thickness
when using a CS-mount or C-mount adapter is 0.050" (1.270 mm). With the adapter removed the maximum filter
thickness is 0.173" (4.4 mm). See Chapter 7 for more information on both types of windows.
Note:
If no glass is installed, dust and debris may collect on the sensor faceplate. Care must be taken when cleaning
a sensor faceplate to avoid damage to the sensor. Please use every precaution to avoid contact with the sensor
faceplate and to keep the camera securely mounted to the optical system or capped when not in use.
Removal of the Window from the Camera
1. If installed, remove any CS-mount or C-mount adapters.
2. Using an SPW606 spanner wrench or equivalent tool carefully remove the SM1 retaining ring and window
as shown below.
Figure 7
Removing the Window
3.6.2. Re-Assembly of the Window
Replacement of the window is done in the reverse order as shown in Section 3.6.1.
Care must be taken when handling the filter. Do NOT over-tighten the retaining ring.