HA0114T Rev 10 May 2019
Chapter 3
3.9 Settings Panel
When the 'Settings' button on the GUI panel is clicked, the 'Settings' window is
displayed. This panel allows data such as scan circle frequency and diameter, and
loop gain to be entered. Note that all of these parameters have programmable
equivalents accessible through the functions on this Control (refer to the
Kinesis API
helpfile
for further details.
3.9.1 Tracking tab
Fig. 3.5 NanoTrak Settings panel - Tracking Tab
Circle
Diameter Mode
- this setting allows the different modes of circle diameter adjustment
to be enabled and disabled.
If '
User
' is selected, the circle diameter remains at the value set by the user in the
User Diameter
parameter (in NanoTrak units). The circle diameter is measured on
a scale such that 10 units is the width and height of the screen. If, for example, the
NanoTrak is driving a 20 micron actuator, a circle diameter of 1 unit will result in a
real diameter of 2 microns. The scanning circle diameter can have values in the
range 0 to 5.0 with a value of 5.0 corresponding to one half the range of piezo
motion.