18.
The measured results will be green if successful and in tolerance. If the measurement failed or the results were
out of tolerance they will be shown in red. Pulling the trigger will move on to measure the next Feature.
19.
If the measurement failed then Review screen is displayed. Tapping on a measured Feature will also show this
screen. Use <Remeasure> to try again, <Delete> to delete an incorrectly taken measurement, <Save> to store
it, <View> to display. <Error code> allows the operator to enter a code for a non-measurable Feature.
20.
When the Group is complete, pull the trigger or tap the top line to return to the Group screen.
If everything is green then the calibration of the GapGun has been verified.
19
18
20
21.
Each Group shows the percentage of Features that have been measured. Tap or Trigger moves to the next
Group. When all Groups are complete Trigger or Back arrow will return to the Jobs screen.
22.
Tap or Trigger moves to the next Checkplan. Trigger or Back arrow will move to the Works Orders screen when
all Checkplans are complete.
23.
“Lock & exit” prepares the results to return them to SPC3D. “Save & exit” saves results if more measurements
are to be made. “Lock & upload” will upload the results to SPC3D immediately.
24.
The WorksOrders shows that the Calibration checkplan is locked ready for uploading. Place the GapGun onto
the Docking Station and tap on the name “Calibration” to begin the upload.
21
22
23
24
Содержание GapGun Pro
Страница 1: ... Third Dimension Confidential Document D13 1604 OP EN 3 3 V7 5 ...
Страница 15: ... Third Dimension Confidential Document D13 1604 OP EN 3 3 V7 5 15 ...
Страница 30: ...30 Third Dimension Confidential Document D13 1604 OP EN 3 3 V7 5 ...
Страница 31: ... Third Dimension Confidential Document D13 1604 OP EN 3 3 V7 5 31 ...
Страница 32: ...32 Third Dimension Confidential Document D13 1604 OP EN 3 3 V7 5 ...