SCPI Command Set
new entry
before
the specified index value or to the end of the list when no index value is
given.
When saved, a new entry is given an index value. Index value starts from 1. When an
index value is specified along with the :SAVE command, the new entry will take the index
of that value and all other following indexes will be incremented by one accordingly, just
as rows in a spreadsheet. Otherwise, the new index will be one up from the index of the
last sweep entry in the list.
When there are no existing entries and an index value other than 1 is specified, an error
will be returned. Similarly for non-existing index location except if the index value is
equal to the value returned by
Syntax
:SWEep:ENTRy:SAVE [integer]
Parameter
[Entry index value]
Input Data Type
Integer
Allowable Values
:COUNt? value + 1
*RST State
N/A
Examples
:SWEEP:ENTR:SAVE
:SWE:ENTR:SAVE 5
:SWEep:ENTRy:ATTenuator
) for the definition of this command.
Examples
:SWEEP:ENTRY:ATTENUATOR ON
:SWEEP:ENTR:ANT?
:SWEep:ENTRy:DECimation
) for the definition of this command.
Examples
:SWEEP:ENTR:DEC 16
:SWEEP:ENTRY:DEC?
:SWEep:ENTRy:FILTer:PRESelect
Refers to the
section (page
) for the definition of this
command.
Examples
:SWEEP:ENTR:FILT:PRES ON
:SWEEP:ENTRY:FILTER:PRES?
:SWEep:ENTRy:FREQuency:CENTer
This command or query defines the center frequency or a range of center frequencies to
sweep. When a range is provided, the sweep will step through the center frequencies
with the value provided by
ThinkRF WSA5000 Wireless Signal Analyzer Programmer's Guide
79