For using frequency output (pin 1) next steps are as following:
IIf. For operation, set FO4, UC1
IIg. Collect data from frequency output. With this setting, you will get wind velocity
corrected by current calibration table and air pressure (so classification and
calibration are applicable). This is without regard to the firmware version
These are the next steps for using serial output:
IIh. For operation set UC1, MS2
IIi. For type *.400 (RS485) collect data v(hPa corr.) from telegram 2 / 4 / 5 / 6 pos. 17 (so
classification and calibration are applicable). This is without regard to the firmware
version
IIj. For type *.401 (MODBUS) collect data v(hPa corr.) from register 30013 / 35013 /
36005 / 36105 / 36205 (so classification and calibration are applicable). This is
without regard to the firmware version
4.3 Determine Occupied Calibration Tables
In order to query the free and the already occupied calibration tables, the
can
be used with parameter 99. In the response string for the command “RC99”, the wind trans-
mitter will output the time stamp of all 10 calibration tables. The time stamps are divided by
the semicolon.
Example:
01RC99\r
2016.04.12;11:12:00;2010.00.00;00:00:00;2010.00.00;00:00:00;2010.00.00;00:00:00;2010.0
0.00;00:00:00;2010.00.00;00:00:00;2010.00.00;00:00:00;2010.00.00;00:00:00;2010.00.00;00
:00:00;2010.00.00;00:00:00;
4.4 Read Calibration Table
Reading the data from the calibration tables is done with the
. The parameter
will indicate the calibration table to be read (0: standard curve, 1 ... 10: Calibration tables). In
the response for the command “RC<parameter>”, the wind transmitter will output all 32 table
entries and calibration certificate number, user-defined text and time stamp of the table.
Example:
01RC1\r
00000;00000;00242;00242;00484;00484;00726;00726;00968;00968;01210;01210;01452;01
452;01694;01694;01936;01936;02178;02178;02420;02420;02662;02662;02904;02904;0314
6;03146;03388;03388;03630;03630;03872;03872;04114;04114;04356;04356;04598;04598;0
4840;04840;05082;05082;05324;05324;05566;05566;05808;05808;06050;06050;06292;062
92;06534;06534;06776;06776;07018;07018;07260;07260;07502;07502;0000001234;Test
;2016.04.12;11:12:00