11.2.15
Command RC
<id>RC<parameter><CR> Read calibration table.
Access:
reading
Description:
With the command, the table indicated in the parameter is read.
The value returned by the command RC depends on the pa-
rameter:
Parameter type:
unsigned integer
Parameter:
Parame-
ter
Description
0 ... 10
Table (0 ... 10)
0: Standard curve
1: Table 1
2: Table 2
...
9: Table 9
10: Table 10
99
Time stamp (date/time) of all tables
(empty tables, i.e. unfilled tables give the following
time stamp: tt;2010.00.00;00:00:00;)
Type return value:
Character string
Return value:
Transfer
parame-
ter
Return value
ASCII representation
Description
1...10
<STX>
sssss[1];iiiii[1];
sssss[2];iiiii[1];
...
sssss[32];iiiii[32];
xxxxxxxxxx;
uuuuuuuuuuuuuuuu;
YYYY.MM.DD;HH:NN:SS
<ETX>
32 sampling points in the table
+ calibration certificate number
+ date/time
sssss: Target value in
0.01[m/s]
iiiii: Actual value in 0.01[m/s]
xxxxxxxxxx: Calibration certifi-
cate number
uuuuuuuuuuuuuuu: User string
with 16 characters
YYYY: Year
MM: Month
DD: Day
HH: Hour
NN: Minute
SS: Second