V2.2  

 

3.

 

Physical Alignment 

 

 
 
 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
           (1)   

 

 

 

 

 

 

 

 

    (2) 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

NOTE:  

To avoid undue stress and strain on the 

cable connection, suitably Tie-wrap them to the 
tripod or the support of choice to reduce potential 
damage.

 

 
 
 
 
 
 
 
 

1.

 

Mount the FSD01 camera (1) on a suitable platform (tripod/ceiling mount)  

2.

 

Mount the ThermaRef 35B1 blackbody (2) on a suitable platform (tripod/ceiling 
mount) at a distance of 1-3m from the FevIR Scan 2 FSD01 camera.  

3.

 

When viewed from behind the FevIR Scan 2 FSD01 camera, the ThermaRef 
35B1 blackbody should be offset to the left or right and not viewed directly in 
line with the front of the field of view.  

4.

 

For optimum system performance the ThermaRef 35B1 blackbody should be 
located at the same point as that used to 

measure subject’s skin temperature.  

This point should be outlined on the floor with tape and people instructed to 
pause momentarily to confirm no alarm is activated. 

5.

 

Tape or barriers should be used to ensure people cannot walk between the 
camera and reference source. 

1-3m 

FevIR Scan 2 

FSD01 camera 

ThermaRef 35B1  

blackbody

 

Содержание FevIR Scan 2

Страница 1: ...not insert any object into the opening or attempt to clean the thermal lens with any material Do not interrupt the movement of the shutter Do not force the shutter with your finger or any other item...

Страница 2: ...SB lead d Network Cable e ThermaRef 35B1 blackbody temperature reference f 24V Power Supply Unit g Laptop Power Supply Unit h Laptop with FevIR Scan 2 software program f e e h a Remove protective cove...

Страница 3: ...ct the ThermaRef 35B1 f to its power supply e and SWITCH ON The ThermaRef is factory set at 35 C and will need approx 20 minutes to stabilise 3 Connect the FSD01 camera b to its power source a DO NOT...

Страница 4: ...m behind the FevIR Scan 2 FSD01 camera the ThermaRef 35B1 blackbody should be offset to the left or right and not viewed directly in line with the front of the field of view 4 For optimum system perfo...

Страница 5: ...witch on and Set Up 1 Power on the FevIR Scan 2 FSD01 camera 2 On the Laptop start the software application Thermoteknix FevIR Scan 2 3 Once running press F7 to enter the Set Up menu 4 Select Setup an...

Страница 6: ...hese can be changed under the various tabs shown below 7 Select the third tab Display 8 Set the Palette to Monochrome for the set up process It can be altered later once everything is up and running 9...

Страница 7: ...to the condition of the system 2 There are three traffic light signals System operating normally Blackbody temperature stable Camera operating normally High temperature detected System NOT operating...

Страница 8: ...l Please also watch video in the manuals folder on desktop Blackbody Thermal Reference Yellow Circle Resizing mp4 1 On Screen you will see a yellow circle Use the laptop mouse pad to grab hold down th...

Страница 9: ...ing on the edge of the circle and changing reducing the shape of the circle 3 Reposition the circle on the image so that it is entirely within the circular hot area of the blackbody 4 The FevIRScan 2...

Страница 10: ...n area Move and resize it with the mouse so it does not include the reference source There is no limit to the area within the field of view and will cover people entering the view of the camera You sh...

Страница 11: ...use to fit inside the hot circular area of the blackbody press F7 to select the setup menu On the Reference Tab change the Diameter number of pixels If necessary then repeat step 3 to reposition the y...

Страница 12: ...V2 2 12 7 Example circle positioning Yellow circle correctly sized and positioned Yellow circle too big Yellow circle too small Yellow circle not aligned and too big...

Страница 13: ...r manual is included on the PC desktop and also on the USB memory stick provided For Technical Support please email support thermoteknix com 1 The Setup menu allows for certain parameters to be change...

Отзывы: