REPLACING BATTERY 
 

WARNING:  

 

Batteries must be changed in Non-hazardous area.  

 

Batteries must be of correct approved type.  

 

Batteries must be installed with correct polarity making sure the (+) end of the battery is aligned with (+) symbol embossed 
in the battery case. 

 

New batteries must not be mixed with old batteries. Batteries must not be mixed with batteries of other manufacturers. 

 

Batteries must not be installed with polarity reversed where one cell could charge another cell.  

 

a) Ensure the instrument is in a non-hazardous area. 
b) Remove the 3 screws holding the front cover on the TP7C or the 2 screws holding the front cover on the TP9. 
c) Remove battery retaining device, push one battery towards the spring contact and lift battery up from the holder, and then remove  
     the remaining battery. 
d) Install each new battery making sure the (+) end of the battery is aligned with (+) symbol embossed in the battery case. 
e)

 

Replace the retaining device and reinstall the cover. 

 
CERTIFIED Batteries for the TP9 and TP7C are as follows: 
 
Manufacturer   

Type 

 

 

Part Number 

 
Duracell  

 

AA (LR6) Alkaline 

MN1500 

Panasonic 

 

AA (LR6) Alkaline 

LR6XWA 

GP (Gold Peak)   

AA (LR6) Alkaline 

GP15A   

 
 
 

MEASURING PROCEDURES 

 
See 

www.thermoprobe.net

 for video on proper use of this instrument. Contact your distributor or ThermoProbe Inc. for names of 

training facilities.  Refer to American Petroleum Institute measurement standard Chapter 7.2. 
 
When the temperature has stabilized both Up and Down arrows will be displayed. 
 
 If you wish to log the temperature for averaging purposes, rapidly press the “On” button twice.  An acknowledgment of a saved 
reading will occur with a display of “LoggEd”.  If the display shows “-------“ then reading was not logged due to a non-stabilized 
temperature. The real time temperature will then be displayed again indicating the next reading is ready to be taken.   
 
If you have completed all the readings needed for your operation, the “f” Button may be used to “LiSt” all the readings logged and 
“At” average of the logged temperatures.  Document your readings and average temperature before powering off the device. 
 
Logged readings and average are not saved when the instrument powers off. (Note: The instrument will automatically power off 20 
minutes after the last button push.)  
 
 
 

AUTHORIZED REPAIR 
 

It is recommended that service beyond the scope of this manual be performed by ThermoProbe, Inc. or one of its authorized 
distributors. 
 
 
 
 
 

 
 

Содержание TP7C

Страница 1: ...ry risks e Power source must be removed before performing any maintenance f Exchange of components other than the batteries may compromise ATEX IECEx certification and shall only be undertaken by Thermoprobe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the tank according to clause 6 3 2 e IEC EN...

Страница 2: ...rdous area to a level below that which will ignite the gases In order to have a fire or explosion fuel oxygen and a source of ignition must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never be great enough to cause ignition TP7C TP9 Hazardou...

Страница 3: ...onic AA LR6 Alkaline LR6XWA GP Gold Peak AA LR6 Alkaline GP15A MEASURING PROCEDURES See www thermoprobe net for video on proper use of this instrument Contact your distributor or ThermoProbe Inc for names of training facilities Refer to American Petroleum Institute measurement standard Chapter 7 2 When the temperature has stabilized both Up and Down arrows will be displayed If you wish to log the ...

Страница 4: ...te Lo bAtt on the display before resuming normal display functions When the batteries are low the backlight will not operate in order to conserve power while the user completes his operations Replace batteries as soon as possible in a safe location after Lo bAtt is noticed this will ensure backlight operation and avoid possible malfunctioning Do not attempt to calibrate the instrument if the Lo bA...

Страница 5: ...ained from ThermoProbe Inc or one of its authorized distributors with part specifications as follows a First follow REPLACING BATTERY instructions a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement White positive sensor wire Black negative sensor wire and sheild c Set the cover and circuit board asi...

Страница 6: ...ndations Calibration must not be performed in any environment considered to be hazardous Equipment needed Ice Bath or other low temperature bath with certified reference thermometer Warm to hot fluid bath between 20 C approx 68 F or higher up to 90 C approx 194 F with certified reference thermometer see Note Optional high temperature oil bath at about 150 C 300 F and certified reference thermomete...

Страница 7: ...ust the device to the actual temperature The On button will decrement the readings and the f Button will increment the reading Once the device temperature matches the actual temperature press the Calibration Button to save the setting The display will scroll SAUE or CAnCEL If the SAUE option is chosen the blinking letter will change to represent the next temperature level A b If the CAnCEL option ...

Отзывы: