CALIBRATION PROCEDURE   

  

 

The calibration mode should only be accessed by qualified personnel with proper equipment; otherwise 
calibration integrity may be compromised. Read the following instructions carefully.   

 

At least 2 points are required to make an adjustment. (2-Point calibration). Additional points can be taken 
(3-Point or 4-Point calibration) to calibrate a large range of temperatures (e.g. 0°F to 300°F) or if you want 
to match specific points in your range. You should include points at the bottom and top of the range. You 
must have the proper equipment for every point of calibration. 

 

Do not attempt to calibrate the instrument if the low battery indicator has been displayed since the new 
calibration values may not be properly stored to memory.   

 

Refer to API 7.2 or another recognized standard for routine calibration verification recommendations.  

 

Calibration must not be performed in any environment considered to be hazardous.   

 

Equipment needed:   

  

 

Ice Bath or other low temperature bath with reference thermometer.   

 

Warm to hot fluid bath between 20°C (approx 68°F), or higher up to 90°C (approx 194°F) with reference 
thermometer. (see Note)*  

 

Optional high temperature oil bath at about 150°C/300°F and reference thermometer.   
 

*Note for limited calibration: If entire range capability of instrument is not required, the 2 point high adjustment 
can be made at a temperature relatively close to the common temperature of the liquid measured and accuracy 
will be maintained within the limited range. For example: If liquid product to be measured is commonly less than 
38°C (approx. 100°F), then a “high point” calibration can be made near that temperature. Temperature accuracy 
above this calibration point cannot be assured.   
 

To calibrate proceed with the following steps:   

 

1.

 

Enter calibration mode by going to 

CALIBRAT 

in the selection Menu. When this is selected, the user will be 

prompted to enter a 3 digit password. Using the “up/down” buttons allows each digit to be adjusted. Once the 
correct value is set, advance to the next digit by pressing the “select” button. After the 3

rd

 digit is set press the 

“select” button to enter Calibration mode. If an incorrect password is entered, 

FAIL

 will display and the unit 

will return to temperature mode. Once in Calibration mode the ‘CAL’ icon on the display will blink. 

The 

calibration mode password is:

 

112

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

2.

 

Select 

ADJUST

 from the calibration menu and then select the desired number of calibration points. Pick 

2PT CAL

3PT CAL

 or 

4PT CAL

 using the select button. 

NOTE:  If the user is not ready to enter the calibration mode, the 

EXIT

 

option can be chosen. 

www.

.com

[email protected]

1.800.561.8187

Содержание TP7-D

Страница 1: ...Thermoprobe Models TP7D and TP9A TP7 D TP9 A USER MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...tified Body Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul Naviglio MI Italy Identification Number 2575 Auditing body QAN Intertek Italia SPA Via Guido Miglioli 2 A 20063 Cernusco sul N...

Страница 3: ...rsed where one cell could charge another cell New batteries must not be mixed with old batteries Batteries must not be mixed with batteries of other models or manufacturers a Ensure the instrument is...

Страница 4: ...up to 10 readings C F Select Celsius or Fahrenheit temperature display DECIMAL Select 0 1 or 0 01 display resolution MEMORY Use up down buttons to display Lowest reading Average reading Highest readin...

Страница 5: ...USB Memory A micro usb connection is available on the circuit board allowing access to calibration logged data Warning Do not access the circuit board in a hazardous location Lowest Highest Average r...

Страница 6: ...a through c to remove batteries b On the circuit board push the terminals clamps down and remove the wires noting the wire lead color code arrangement See Figure 1 White positive sensor wire Green cab...

Страница 7: ...ence thermometer Note for limited calibration If entire range capability of instrument is not required the 2 point high adjustment can be made at a temperature relatively close to the common temperatu...

Страница 8: ...return the unit to a factory calibration curve calibration adjustment will still be required o The RESETCAL reset to factory calibration feature should be used if the probe is replaced using a 2 wire...

Страница 9: ...nd shall only be undertaken by ThermoProbe or one of its qualified service providers See also Authorized Repair section g To reduce the risk of fire or explosion this device must be bonded to the vess...

Страница 10: ...must be present An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere but the system is designed so the electrical energy or thermal energy of a particular instrument...